Before initiating the cleaning process, make sure the material to be cleaned may be submitted to ultrasonic
cleaning (cavitation).
Never put your hand or any other part of your body inside the cleaner when it is in use, this procedure may be harmful to
your health.
5
INSTALLATION INSTRUCTIONS
1
2
.
The storage must be done in a place protected from the weather action (indoors) in normal temperature conditions
Install the
at
ultrasonic cleaner in an exclusive sterilization room, ventilated and clean, on a flat, leveled and firm surface,
a proper height for the operator to handle it Leave enough room next to the equipment for handling the instruments/articles.
.
Table 1
3
The voltage of the ultrasonic cleaner is 127V
or
220V, as informed on the identification label locat
ed on the bottom
of the equipment. (Fig. 1, pág. 6).
Make sure the wiring
where the equipment will be installed is in
voltage
accordance with these values.
For the installation, use a three pin outlet with grounding pin
(2P + T, 10A).
Never
use extensions, adapters or voltage transformers
.
ATTENTION
Grounding is very important for the safety of the operator and the warranty of your equipment
!
. Install the
ultrasonic cleaner where it can be easily unplugged
.
Electrical
See “
Authorized Technical Assistance
” (Page 13). In case the user chooses an unauthorized electrician or technician to
install the equipment, Cristófoli Biossegurança will not be responsible for any services performed in disagreement with the
instructions described in the product manual.
The electrical installation must follow the data from the table below.
MODEL
AMPERAGE
WIRING GAUGE (DIAMETER)
Cristófoli Ultrasonic
Cleaner (127 V)
1,5 A
Cristófoli Ultrasonic
Cleaner (220 V)
The wiring from the breaker to the electrical outlet
used to connect the equipment
must have a diameter of at least 2.5 mm.
0,8 A