background image

CRISTÓFOLI

U L T R A S O N I C   C L E A N E R

I N S T R U C T I O N    M A N U A L

7

DEGASSING / USING THE BEAKER

 

The water degassing has the purpose of removing the gases dissolved in the 

water that impair the cavitation process and must be repeated every time the 
water/solution is changed. The procedure consists in leaving the equipment on 
for 8 minutes (480s cycle) without any instruments /articles in the vat.

 

When using the beaker (accessory that must be purchased separately used 

to clean very small items (burs, files, etc.) that would certainly go through the 
basket orifices falling on the bottom of the vat, remember that the water inside 
the beaker must also be degassed  Use always a heat resistant glass beaker 

.

specific for this purpose. Put the water in the stainless steel vat, the basket, the 
beaker on top of the basket and water inside the beaker, seal it with some PVC 
film (the same used for food wrapping) and proceed the degassing.

 

Always use the lid when the equipment is operating with the basket, when 

using the beaker, remove it, unless the beaker being used has a compatible 
size for the basket depth and allow the use of the lid.

 

The Cristófoli Ultrasonic Cleaner may also be used to clean dental prosthesis 

like dentures, bridges or orthodontic devices like braces, etc., as shown by the 
drawing (Fig. 6).

Fig.6

Fig.7

FUSE  - 

Internal  safety  device  built  into  the  electronic  board  which

.  See 

  cuts  the  power  from  the  equipment

Troubleshooting

 

(Page 12).

OVERHEATING PROTECTION SENSOR

It protects the cleaner against overheating. If the 

lig

 (On) key is pressed, the 

 - 

display  shows  the  remaining  time  and  there’s  no  cavitation  (production  of  ultrasonic  waves),  that  means  that  the 
equipment  needs  to  cool  down  for  about  15  minutes.  After  that,  press 

lig

  again  and  the  equipment  will  resume 

operation.

1

SAFETY DEVICES

2

 

For a deeper cleaning, with the presence of fat, optionally, 

the water heating function can be used. Just press the 

AQ

 key to 

start the water heating (Fig. 7) that will take approximately 40 
minutes to achieve the ideal temperature for cleaning, which 
is 65 ºC.

 

Once the right temperature is reached, heating will turn off 

automatically The red LED on the left side of the display will be 

on  when  the  heating  is  in  use If  it’s  needed  to  reduce  the 

heating  time,  water already  warm 

 

may  be  used  by  the 

operator. Do not use water too hot (above 65ºC), this will hinder 
the cleaning of the instruments/articles.

HEATING

Red

LED

AQ

DES

LIG

MD

Summary of Contents for Ultrasonic cleaner

Page 1: ...op innovative solutions to protect life and promote health Crist foli Equipamentos de Biosseguran a Ltda established at Rodovia BR 158 n 127 Jardim Curitiba in Campo Mour o Paran Brasil manufactures b...

Page 2: ...g the Beaker Heating How to Use the Crist foli Ultrasonic Cleaner T echnical Data How to Identify your Ultrasonic Cleaner Preventive Maintenance Troubleshooting Quality Control Authorized Technical As...

Page 3: ...ate of manufacture Good Manufacturing Practices 5 P B F LOT SN Fragile Handle with care Dangerous Electrical Tension Serial Number Batch code Keep away from sunlight Recyclable Keep dry ISO 13485 ISO...

Page 4: ...BR 158 n 127 Campo Mour o PR Brasil CEP 87309 650 CNPJ 01 177 248 0001 95 Inscr Est 90104860 65 Website www cristofoli com e mail cristofoli cristofoli com Responsible Technician Eder William Costa C...

Page 5: ...he bottom of the equipment Fig 1 p g 6 Make sure the wiring where the equipment will be installed is in voltage accordance with these values For the installation use a three pin outlet with grounding...

Page 6: ...or Front View Fig 1 Off key On key Mode Key Adjustment of operation times Green LED indicates that the equipment is plugged in or in operation Right red LED indicates that the overheating protection i...

Page 7: ...llow the use of the lid The Crist foli Ultrasonic Cleaner may also be used to clean dental prosthesis like dentures bridges or orthodontic devices like braces etc as shown by the drawing Fig 6 Fig 6 F...

Page 8: ...ing to the recommendations specified by its manufacturer We suggest for your convenience that you use a small medicine measuring cup or syringe without the needle to measure the volume of detergent ma...

Page 9: ...the lig key to start cleaning Fig 14 at this moment the operator will hear the typical sound of the cavitation process key Fig 15 To pause the cycle press the des to resume cleaning press lig again 6...

Page 10: ...working ambient temperature Certifications 2 5 liters 2 1 liters 2 5 kg with components 3 kg in the box 28 5 x 18 5 x 22 5 cm 28 8 x 14 8 x 8 cm 127 or 220 V 50 60 Hz 160 Watts 127 V model 170 Watts...

Page 11: ...teel vat and change the cleaning solution after each cycle to maintain the cleaning efficiency The contaminated solution must be disposed according to the instructions of the topic How to Use the Cris...

Page 12: ...ssary arrangements for the electrical wiring are made as shown on item Installation Instructions Page 5 See Authorized Technical Assistance Page 13 Plug the power cable on both ends the power connecto...

Page 13: ...vities All products have a useful life span but it is not possible to determine how long as it varies according to the intensity and how the equipment is used or handled Crist foli Equipamentos de Bio...

Page 14: ...s dents drops short circuits fire etc Sinister theft or robbery Improper use installation of the equipment and or plugging it into an incorrect voltage outlet Problems due to failure of the power supp...