42
© 2020 All rights reserved. Data subject to change without notice.
cGas Detector for Indoor Public Spaces - Installation Manual
Rev. A | 2020.11
From the main display, press ENTER.
Enter passcode 3032 using the UP button and ENTER button.
Enter Password
3032
Press ENTER to access the Calibrate Menu.
Calibrate Menu
Press Enter
In the Selected Channel menu, if Temperature is not displayed, press ENTER
and use the UP button to scroll t
o fi nd Temperature. Press ENTER.
Selected Channel
CO
Selected Channel
>Temperature
Use the UP button to fi nd Temperature Unit. If you want to change the unit type,
press ENTER.
Temperature Unit
Celsius
Press the UP button to choose Fahrenheit press ENTER and ENTER to exit.
Temperature Unit
>Fahrenheit
7.5 Temperature and / or Relative Humidity Offset
NOTE:
This menu item only applies if the cGas Detector has the -RHT option
installed.
Depending on the confi guration, the device will show the temperature in either
Celsius or Fahrenheit. The units can be changed at any time, refer to
Section 7.4
Change Units (
o
C or
o
F) of Temperature Readings
.
The temperature and relative humidity sensors come pre-calibrated from the
factory. If you fi nd that the readings on the cGas Detector are higher or lower
than another measurement device, you can adjust the reading by setting an
offset value so the reading is more accurate. The Temperature offset value is a