- 19 -
CALCulate:LIMit:UPPer?
Returns the upper measurement limit in Ohms.
CALCulate:LIMit:STATe <ON|OFF> or <1|0>
Enables or disables the measurement limit facility.
CALCulate:LIMit:STATe?
Returns “0” (OFF) or “1” (ON).
CALCulate:LIMit:ALARm <ON|OFF> or <1|0>
Enables or disables the 'failed limit’ audible alarm.
CALCulate:LIMit:ALARm?
Returns “0” (OFF) or “1” (ON).
4.5 Other function Commands
Temperature Compensation Group
SENSe:TCOMpensate:STATe <ON|OFF> or <1|0>
Enables or disables the temperature compensation facility.
Temperature compensation is not available in FAST mode. Attempting to set the
state “ON” will generate an “execution error”.
SENSe:TCOMpensate:STATe?
Returns “0” (OFF) or “1” (ON).
SENSe:TCOMpensate:MODE <EXT> or <MAN,
manual temperature
>
Sets the temperature of the test-piece to either an external Pt100 probe
measurement, or to a fixed, manually-entered temperature, the unit of which is
the current temperature unit (ºC or ºF). Range is 0 - 100 ºC (32 - 212°F)
SENSe:TCOMpensate:MODE?
Returns “EXT”, or “MAN”,<
manual temperature
>.
SENSe:TCOMpensate:REFerence <
temperature value
>
Sets the reference temperature. This is the temperature to which resistance
measurements are compensated (normally 20ºC). It is entered in the current
temperature unit (ºC or ºF). Range is 0 - 50ºC (32 - 122°F), in integer values of
ºC, or the ºF equivalent.
SENSe:TCOMpensate:REFerence?
Returns the reference temperature in the current temperature unit.
SENSe:TCOMpensate:COEFficient <CU|AL>or <USER,
user coefficient
>
Sets the temperature coefficient to Copper, Aluminium or to a user-defined value,
which is entered in ppm/ºC (0 - 9999).
SENSe:TCOMpensate:COEFficient?
Returns “CU”,<fixed copper coefficient>, or “AL”,<fixed aluminium coefficient>, or
“USER”,<user coefficient>. The coefficients are quoted in ppm/ºC.
UNIT:TEMPerature <C|CEL|F|FAR>
Sets the temperature unit to ºCelsius or ºFarenheit.
UNIT:TEMPerature?
Returns “CEL” or “FAR”.