Remote manual triggering
As an alternative to pressing the MEASURE/STOP key to trigger, a measurement, the remote
start facility can be used. Connect a switch between pins 1 and 5 of the RS232 input
connector and close the contacts to start a measurement.
HINT:
Connecting a footswitch or measuring probe with an external start button permits single
person operation. Values can be stored in the datalog for future analysis.
Automatic Measurement Mode
In this trigger mode the system allows a measurement to be taken simply by connecting the
measurement leads to the test piece. The measurement will not be triggered if the display is
in the menu mode.
Automatic measurement in single mode:
In most modes (Simple, Limits, Filter, Settling, Autorange, Temp Comp, any current mode,
and normal Datalog), Auto trigger causes one complete measurement to be made and held
on the display.
When in Datalog Timer mode:
Auto trigger starts to take the prescribed number of measurements at the selected timer
intervals.
HINT
: Removing and replacing the measurement leads will have no further effect, other than
causing an error indication.
Automatic measurement in continuous mode:
In most modes (Simple, Limits, Filter, Settling, Autorange, Temp Comp, any current mode,
and normal Datalog):
Auto trigger starts continuous measurements.
HINT:
Removing and replacing the measurement leads will have no further effect, other than
causing an error indication.
Automatic measurement when in Datalog Timer mode:
Auto trigger starts to take the prescribed number of measurements at the selected timer
intervals, maintaining the current at all times.
Hint:
removing and replacing the measurement leads will have no further effect, other than
causing an error indication.
4.9.4
LIMITS
DO7PLUS RS-232 CONNECTION SOCKET
Pins numbered as viewed from rear of plug (cable side)
x
PIN 1 ... REMOTE START
x
PIN 5 ...GROUND
1
2
3
4 5
6
7
8
INSTRUMENT MENU
LIMITS:
6.0000 k
5.0000 k
OFF
S
ON
MAX
MIN
ESC
OK
MAX
MAX LIMITS: k
(0 -- 6k
)
ESC
UNIT CLEAR
OK
MIN LIMITS: k
(0 -- 6k
)
ESC UNIT
CLEAR
OK
Operating Instructions
Digital Microhmmeter Type DO7 PLUS
22