6 CALIBRATION
The DO7PLUS is designed with ease of calibration in mind and comes complete with a
manufacturer’s calibration certificate. A UKAS calibration certificate is also available and
should be ordered with the instrument if required. The calibration should be verified annually
and if necessary adjusted. The unit may be returned to CROPICO who offer a full calibration
service.
For those who wish to perform their own calibration, enter the calibration mode by pressing
F4 (SET UP) scroll down the options and highlight CAL in the menu.
HINT
: To navigate around the screen use the navigation arrows
Where an item is highlighted with an arrow
This indicates that additional options are available for that parameter and the right
navigation arrow should be pressed to reveal those options.
The display will prompt you to enter a PASSCODE. The factory default passcode is 9252.
Press F5 (OK) once the passcode is entered and the last calibration date and time will be
displayed.
BACKLIGHT
SETUP MENU
BEEPER ON
DATE FORMAT
DD/MM/YY
DATE
29/04/08
TIME
16:23
LANG
ENG
LINE FREQ
50HZ
TEMP UNIT
G
C
REMOTE
9600
VERSION VER
1.0
PASSCODE
CAL
EXIT
CAL
CAL
Operating Instructions
Digital Microhmmeter Type DO7 PLUS
35