background image

CROPICO LTD – DO7

Issue 7/2001

9

10. POWER 

UP

When the DO7 is first switched on, a lamp test is automatically performed.  All display segments
are illuminated followed by each LED lighting in sequence.  NOTE  the LOCK LED is not fitted and
will therefore will NOT light. The microprocessor checks for correct internal operation and indicates
"P" if all tests pass O.K.  At the same time the software version will be displayed and held for
approximately 2 seconds.  The display will typically read "P 1.1".  Should the internal checks
indicate an error, then the display will read "HELP".  Contact our Service Department or your local
agent for assistance.

The DO7 will perform an automatic zero sequence and finally sets to the following default start-up
mode ready for use.  The selected measuring range will be 60

, and the DO7 will then be in

stand-by mode.

11. MEASUREMENT

Connect the resistance to be measured (Rx) to the measuring terminals in accordance with the
diagram on the instrument panel.  Select the range required, the measurement mode, ie. +, - or
average.  The LED lamp will light to indicate which buttons are active.  To initialise the
measurement, press and hold the ON button, the measurement will cease once this button is
released and the instrument will return to the STANDBY state.  Should you wish to initialise a
continuous reading, then press the LOCK + ON buttons.  The measurement will continue until the
STANDBY button is pressed or the ON button pressed again. An audible BLEEP will sound during
this time to warn the operator that measurement is still in progress, and up to 10 Amps measuring
current is draining the batteries.

DISPLAY HOLD FUNCTION

This works as follows:-

1. When returning to the STANDBY condition the LAST measured value will remain on

DISPLAY for 30 SECONDS and then blank.

This occurs when:

Pressing the STBY key, or
 Releasing the momentary ON key, or
 Releasing the momentary external footswitch.

2.  Pressing any key (except LOCK) during the 30 second period will force the display to

blank and allow normal operation to resume.  It is recommended that the STBY key be
pressed to carry out this function.

Over-Range

The display will indicate - - - -.
Select a higher range.

Open Circuit Lead

O/C LEAD will be displayed and the display will indicate - - - - if the instrument detects that the
lead resistance is too high.  The C terminals are checked for compliance voltage.  Measurement
should not be made if this warning message is displayed.  This warning will also be displayed if the
internal protection fuse is open-circuit. When in STANDBY mode this LED will always be lit.

Connections

When making good quality measurements, it is important to ensure that all measuring leads are in
good condition, and less than 0.2 ohms resistance.

Summary of Contents for DO7

Page 1: ...DIGITAL MICROHMMETER TYPE DO7...

Page 2: ...T 5 7 TECHNICAL DATA 5 8 DESCRIPTION OF CONTROLS 6 Range Selection 7 Error Status Lamps 7 9 METHODS OF MEASUREMENT 7 10 POWER UP 9 11 MEASUREMENT 9 Over Range 9 12 REMOTE 10 13 PRINTER OUTPUT 10 14 PR...

Page 3: ...the apparatus make sure that it is set to the voltage of the power supply The mains plug shall only be inserted in a socket outlet provided with protective earth contact The protective action must no...

Page 4: ...bility is achieved by using our own manufactured internal resistance standards 4 CASE DESIGN The case is ruggedly constructed from an ABS polycarbonate alloy coloured safety yellow A strong internal s...

Page 5: ...use The batteries will recharge in approximately six hours and the DO7 lid should be open to allow for maximum ventilation IMPORTANT Always connect instrument to mains supply after use and top up batt...

Page 6: ...ing currents Measuring time Approx 0 5 seconds Polarity Forward and reverse measurement current may be selected plus average mode which automatically displays the average value of positive and negativ...

Page 7: ...DO7 Issue 7 2001 6 8 DESCRIPTION OF CONTROLS 4 Digit LCD Display Range Switches On Off Power Switch Calibration Measuring Terminals RS232 Printer Socket Battery Status Indicator Mains Socket Remote S...

Page 8: ...supply is connected to the instrument CAL Calibration mode has been initialised by the key switch and the instrument is in calibration mode O C One of the measuring leads is open circuit LEAD too high...

Page 9: ...CROPICO LTD DO7 Issue 7 2001 8 Fig 9 1 Combined current and potential probes Fig 9 2 Various types of resistors...

Page 10: ...NDBY state Should you wish to initialise a continuous reading then press the LOCK ON buttons The measurement will continue until the STANDBY button is pressed or the ON button pressed again An audible...

Page 11: ...he last value measured when switching from ON to STBY This key toggles ON OFF Baud Rate This key will display the currently set value when pressed To change the value press CLE to clear and then use t...

Page 12: ...ENT REQUIRED Cropico Calibration Standard Type MTS2 Set of 2 low thermal emf leads Set of 2 general purpose leads 10A rating Calibration key no 901 17 PREPARATION 17 1 Ensure all retaining screws are...

Page 13: ...FICATION 19 1 Ensure that the panel switch is set to RUN and the calibration key is removed 19 2 Set the calibration rig to 40 Ohm Zero 19 3 Set the DO7 to 60 Ohm range ve and press LOCK ON Wait for t...

Page 14: ...ay be used in place of the Cropico Calibration Standard The standards must have an uncertainty better than 0 01 if the full calibration accuracy is to be achieved Full consideration of the DO7 measuri...

Page 15: ...CROPICO LTD DO7 Issue 7 2001 14...

Page 16: ...has been detected This test is performed at power up and failure is indicated by the HELP message 21 2 Turn the DO7 power ON Wait for the display test to complete and the HELP message to appear 21 3...

Page 17: ...CROPICO LTD DO7 Issue 7 2001 16...

Page 18: ...e leads HSO2 RS Duplex Handspikes as HSO2 but with remote start button fitted LSO3 Large Kelvin clips with 3 metre leads accept cables and bars up to 38mm diameter LSO4 As LSO3 but with 3 metre leads...

Page 19: ...S THERMOCOUPLE REFERENCE JUNCTIONS THERMOCOUPLE SELECTOR SWITCHES PT100 SIMULATORS PRECISION WATTMETERS INSULATION TEST SETS UNIVERSAL CALIBRATORS POWER ANALYSERS HIPOT TESTERS ELECTRICAL SAFETY TESTE...

Reviews: