background image

CROPICO LTD – DO7

Issue 7/2001

13

19.8

Set the calibration rig to 400m Ohm, +FS and wait for the reading to settle.

The reading should be 0.40 +/-1 digit.  Note the actual reading in the Calibration Test
Results sheet.

19.9

Set the calibration rig to 40m Ohm, +FS and wait for the reading to settle.

The reading should be 0.04 +/-1 digit.  Note the actual reading in the Calibration Test
Results sheet.

19.10

Repeat steps 5.2 thru' 5.6 for the remaining ranges.

19.11

Turn the DO7 power OFF and remove the test leads.

19.12

Affix a calibration seal label to the instrument.

19.13

Sign and date the Calibration Test Results sheet.

* *  END OF NORMAL CALIBRATION PROCEDURE  * *

20.  USING DISCRETE RESISTANCE STANDARDS

20.1

Discrete, 4-terminal resistance standards may be used in place of the Cropico Calibration
Standard. The standards must have an uncertainty better than 0.01% if the full calibration
accuracy is to be achieved.  Full consideration of the DO7 measuring current must also be
taken into account when selecting  the standards.  The DO7 can be calibrated to standards
between 1000 and 6000 digits for each range using the KYB entry.  However, full accuracy
will only be achieved if standards of 4000 to 6000 digits are used.  The following nominal
standards are recommended:

40,   4,   400m,   40m,   4m   and   400u Ohm

20.2

In all measurements the P1 and P2 terminals are conventionally connected.  Bear in mind
that thermal emfs may be present.  The DO7 is particularly sensitive to these on the 600u
Ohm range.

20.3

For FS measurements, connect the C1 and C2 leads in the conventional manner (see Fig
20.1).

20.4

For ZERO measurements, connect the C1 and C2 leads to the C2 terminal of the standard
(see Fig 20.2).

Summary of Contents for DO7

Page 1: ...DIGITAL MICROHMMETER TYPE DO7...

Page 2: ...T 5 7 TECHNICAL DATA 5 8 DESCRIPTION OF CONTROLS 6 Range Selection 7 Error Status Lamps 7 9 METHODS OF MEASUREMENT 7 10 POWER UP 9 11 MEASUREMENT 9 Over Range 9 12 REMOTE 10 13 PRINTER OUTPUT 10 14 PR...

Page 3: ...the apparatus make sure that it is set to the voltage of the power supply The mains plug shall only be inserted in a socket outlet provided with protective earth contact The protective action must no...

Page 4: ...bility is achieved by using our own manufactured internal resistance standards 4 CASE DESIGN The case is ruggedly constructed from an ABS polycarbonate alloy coloured safety yellow A strong internal s...

Page 5: ...use The batteries will recharge in approximately six hours and the DO7 lid should be open to allow for maximum ventilation IMPORTANT Always connect instrument to mains supply after use and top up batt...

Page 6: ...ing currents Measuring time Approx 0 5 seconds Polarity Forward and reverse measurement current may be selected plus average mode which automatically displays the average value of positive and negativ...

Page 7: ...DO7 Issue 7 2001 6 8 DESCRIPTION OF CONTROLS 4 Digit LCD Display Range Switches On Off Power Switch Calibration Measuring Terminals RS232 Printer Socket Battery Status Indicator Mains Socket Remote S...

Page 8: ...supply is connected to the instrument CAL Calibration mode has been initialised by the key switch and the instrument is in calibration mode O C One of the measuring leads is open circuit LEAD too high...

Page 9: ...CROPICO LTD DO7 Issue 7 2001 8 Fig 9 1 Combined current and potential probes Fig 9 2 Various types of resistors...

Page 10: ...NDBY state Should you wish to initialise a continuous reading then press the LOCK ON buttons The measurement will continue until the STANDBY button is pressed or the ON button pressed again An audible...

Page 11: ...he last value measured when switching from ON to STBY This key toggles ON OFF Baud Rate This key will display the currently set value when pressed To change the value press CLE to clear and then use t...

Page 12: ...ENT REQUIRED Cropico Calibration Standard Type MTS2 Set of 2 low thermal emf leads Set of 2 general purpose leads 10A rating Calibration key no 901 17 PREPARATION 17 1 Ensure all retaining screws are...

Page 13: ...FICATION 19 1 Ensure that the panel switch is set to RUN and the calibration key is removed 19 2 Set the calibration rig to 40 Ohm Zero 19 3 Set the DO7 to 60 Ohm range ve and press LOCK ON Wait for t...

Page 14: ...ay be used in place of the Cropico Calibration Standard The standards must have an uncertainty better than 0 01 if the full calibration accuracy is to be achieved Full consideration of the DO7 measuri...

Page 15: ...CROPICO LTD DO7 Issue 7 2001 14...

Page 16: ...has been detected This test is performed at power up and failure is indicated by the HELP message 21 2 Turn the DO7 power ON Wait for the display test to complete and the HELP message to appear 21 3...

Page 17: ...CROPICO LTD DO7 Issue 7 2001 16...

Page 18: ...e leads HSO2 RS Duplex Handspikes as HSO2 but with remote start button fitted LSO3 Large Kelvin clips with 3 metre leads accept cables and bars up to 38mm diameter LSO4 As LSO3 but with 3 metre leads...

Page 19: ...S THERMOCOUPLE REFERENCE JUNCTIONS THERMOCOUPLE SELECTOR SWITCHES PT100 SIMULATORS PRECISION WATTMETERS INSULATION TEST SETS UNIVERSAL CALIBRATORS POWER ANALYSERS HIPOT TESTERS ELECTRICAL SAFETY TESTE...

Reviews: