Page 11 of 14
- On the dosimeter DR020, press the "T2: Print Memory" function.
The information is printed in the following format:
No. address dose
App.factor Foil
factor
001
5D64
048.1
100
100
002
5D68
050.5
100
100
003
5D6C
051.7
100
100
004
5D70
052.7
100
100
005
5D74
055.2
100
100
- Select
Save as
from the
File
menu and place the file into the appropriate directory. This file
can imported into a spreadsheet program, for example, for further processing.
11. Servicing / malfunctions
The device has three different modes for error detection and servicing.
- Service mode 1
automatic activation
- Service mode 2
manual activation
- Service mode 3
automatic activation
11.1 Service mode 1
If the device can no longer calibrate "Error T2: service program" is displayed. This function can
be forced by placing a piece of paper into the measuring head during calibration.
In this mode,
- red on
(> 500)
- off
(0)
- green on
(> 500)
can be controlled separately. Check each displayed value which represents the intensity. This
value must be above 500 for red and green with an empty measuring head. A value below 500
indicates that the device must be readjusted or is defective. These intensity values can also be
displayed in the main menu (see section 4.6.2).