11.2 Service mode 2
If the keys T1 and T2 are pressed at the same time during start-up, the device is switched to
service mode 2. This mode displays important data to simplify diagnostics.
The instrument-specific calibration factor (App.fact) which compensates for the hardware of
different devices can also be set in this mode. Do not change the instrument-specific calibration
factor! Adaptations regarding different dosimeter film batches can be made by modifying the
film-specific calibration factor (foil fact).
Flow chart / menu structure