1.  Purpose of the device 

 
This dosimeter is used to measure ionizing radiation in processes such as cross-linking of paints or 
plastics. 
As a measurement instrument, it is extremely simple to use: a measuring film is exposed to radiation 
and then evaluated using the dosimeter in a matter of seconds. The measured value is displayed in 
kGy. By offsetting optical components, performing measurements using a 10-Bit A/D converter and 
averaging values over several measurements, the dosimeter offers a very high degree of accuracy. 
Measurement errors only occur if there are irregularities in the measuring film (thickness and color 
distribution). Thanks to these features, the dosimeter is a valuable tool in quality assurance. 
 
 
1.1 

Electron beam irradiation 
When using electron beam equipments a three-dimensional determination of dose distribution is 
a must for reproducible production procedures. 

 

In the case of low-energy electron beam equipment, the short range of the electrons causes 
considerable alteration in the absorbed dose with the penetration depth. One useful dosimeter 
employs polyvinyl butyral (PVB) with pararosaniline cyanide: this colors on irradiation with 
electrons or UV (red discoloration). 
This very thin film (approx. 20 µm or 20 g/m

2

) has good resolution in depth doses for 

measurements with low-energy electrons. As the film is supplied by the meter, it is also possible 
to measure the dose distribution across the effective width at high resolution, as follows: 
Radiation is directed onto a strip which corresponds to the beam width and evaluation then 
performed at cm intervals, for example. 

 
 

2. Device 

design 

 

The dosimeter is accommodated in a sealed high-grade steel housing which effectively shields 
it from interference. A keyboard overlay on the function keys and the display protects against 
dirt. 
To allow a better view while working, you can tilt the device at an angle using the retractable 
stands. Just three keys are necessary to activate all of the device functions; all necessary data 
is shown on a 2 x 20 character display screen.  
The measuring head, which has a measurement slot 0.5 mm in width, contains the transmitter 
and receiver optical components. 
A processor the size of a check card is located inside the device: this controls all of the 
functions and dose calculation. 
 
 

3. Power 

supply 

 

The power is supplied by a standard power supply unit. The measurement device is designed 
for continuous operation. The device can be switched off with the power switch located on the 
device. 
 
The serial interface can be used to read the stored values into a PC for further processing. 

 

Summary of Contents for DR 020

Page 1: ...0 281 SE 302 44 Halmstad Telefax DE 72147 Nehren Telefax 46 0 35 14 82 06 49 0 7473 920 282 EB_DOS_01_Manual_EN_01 doc Page 1 of 14 Operating manual for dosimeter type DR020 1 Purpose of the device 2 Design 3 Connection 4 Operation 5 Calibration 6 Application 7 Method of measurement 8 Principle of measurement 9 Measuring film 10 Measured values via the interface 11 Malfunctions servicing 12 Techni...

Page 2: ...n or use of this document is not permitted without express written authority Electron Crosslinking AB shall have no liability for any indirect or speculative damages arising from the use of this product Technical data subject to change Other product and corporate names may be trademarks of other companies and are only for explanation and to the owner s benefit without intent to infringe ...

Page 3: ...coloration This very thin film approx 20 µm or 20 g m2 has good resolution in depth doses for measurements with low energy electrons As the film is supplied by the meter it is also possible to measure the dose distribution across the effective width at high resolution as follows Radiation is directed onto a strip which corresponds to the beam width and evaluation then performed at cm intervals for...

Page 4: ...t you should switch on the device 30 min before use and then recalibrate it after the half hour In this way accurate measurements are achieved Once calibration is complete the device is ready to perform measurements When a radiated film see item 10 is inserted a dose reading immediately appears on the display A constant measured mean value is displayed after 8 to 32 measurements Constant values ar...

Page 5: ...roceed to the next menu 4 2 1 T2 Print memory See section 10 Measured values via the interface 4 2 3 T3 Ready The T3 key is used to return to menu 1 4 3 Menu 3 Clear memory 4 3 1 T1 Next The T1 key is used to proceed to the next menu 4 3 2 T2 Clear memory The T2 key is used to erase all data from the memory Subsequently the message memory is empty is displayed and you are returned to menu 1 4 3 3 ...

Page 6: ...e above 500 If this value is not reached the device needs to be readjusted 4 6 3 T3 Ready The T3 key is used to return to menu 1 5 Calibration After starting the device is calibrated After prolonged operation recalibration of the device is recommended to compensate for any thermal drifts of the optical components for example thereby ensuring accuracy of the dose evaluation 5 1 Measurements with a ...

Page 7: ...on This method compensates for variations in the thickness 2 3 µm of the dosimeter measuring film and yields improved resolution A resolution of 10 g m2 is achieved with a correspondingly accurate dose at the end of the stack via the measuring film of 20 g m2 At 20 g m2 dose loss is minimal thus reducing the incidence of measurement errors Typical depth dose with a resolution of 20 g m2 6 2 Measur...

Page 8: ...he other color is reset until both of the measured intensities are roughly equivalent These intensities are stored in the memory as the values Ig0 and Ir0 By comparing the difference between the intensities scattered light are suppressed The measured values are evened out by software components and a mean value calculated using 8 of the evened values A measurement period of 3 times 20ms is require...

Page 9: ...ng of the dosimeter film Please note that the dosimeter film must always lie on top when measuring doses with low energy electrons The dosimeter film should be evaluated without the supporting film If you wish to perform evaluation using supporting film the dosimeter must first be calibrated with the supporting film and without the dosimeter film The maximum operating temperature must be taken int...

Page 10: ...using COM 1 or COM 2 Carry out the following steps to print the data Start the Windows program HyperTerminal From the File menu select New Connection and confirm your selection Enter the name of the file into the New Connection window e g MeasuredValues01 and confirm the name In the File name window e g MeasuredValues01 select Connect Using choose Direct to COM 1 or COM 2 or your interface and con...

Page 11: ...modes for error detection and servicing Service mode 1 automatic activation Service mode 2 manual activation Service mode 3 automatic activation 11 1 Service mode 1 If the device can no longer calibrate Error T2 service program is displayed This function can be forced by placing a piece of paper into the measuring head during calibration In this mode red on 500 off 0 green on 500 can be controlled...

Page 12: ...ata to simplify diagnostics The instrument specific calibration factor App fact which compensates for the hardware of different devices can also be set in this mode Do not change the instrument specific calibration factor Adaptations regarding different dosimeter film batches can be made by modifying the film specific calibration factor foil fact Flow chart menu structure ...

Page 13: ... fact instrument specific calibration factor T2 calls up the menu which can be used to set the instrument specific calibration factor Any factor between 0 255 can be set This allows differences in components to be compensated for thereby ensuring identical dose readings for different devices This factor is retained when exiting with T3 and is not to be modified The device must be calibrated see se...

Page 14: ...er supply socket center contact housing Current 370 mA Keyboard 3 mechanical keys covered with a keyboard overlay Display Dot matrix display 20 characters 2 lines Optical transmitter Transmitter diode wavelength red 635 nm green 565 nm Optical receiver Receiver diode wavelength 500 700 nm Interface Serial interface Data output via 0 modem cable Electronics Check card sized processor with A D conve...

Reviews: