Page 5 of 14
4.1 Menu 1 / main menu
Dose reading, Save, Calibration
4.1.1 T1 Next
The T1 key is used to proceed to the next menu. Even though this function does not appear on
the display, this key is active.
4.1.2 T2: Save
The T2 key is used to save the value that is currently displayed. The count value of the value
(up to 8128 values can be stored) is displayed when the value is stored.
4.1.3 T3: Calibration
The device can be recalibarated after the turn-on time (in accordance with the procedure
described in section 4) and while the device is in operation.
4.2 Menu 2
memory
4.2.1 T1 Next
The T1 key is used to proceed to the next menu.
4.2.1 T2: Print memory
See section 10 "Measured values via the interface"
4.2.3 T3: Ready
The T3 key is used to return to menu 1.
4.3 Menu 3
Clear
memory
4.3.1 T1 Next
The T1 key is used to proceed to the next menu.
4.3.2 T2: Clear memory
The T2 key is used to erase all data from the memory. Subsequently the message „memory is
empty“ is displayed and you are returned to menu 1.
4.3.3 T3: Ready
The T3 key is used to return to menu 1.
4.4 Menu 4
Change foil factor
4.4.1 T1 Next
The T1 key is used to proceed to the next menu.
4.4.2 T2: Change foil factor
Every new dosimeter film batch may have a different offset. This can be corrected by entering a
film-specific value. The keys T1+ and T2- can be used to select a factor in the range of 0 to 255.
4.4.3 T3: Ready
The T3 key is used to return to menu 1.