Page 6 of 14
4.5 Menu 5
Choose
language
4.5.1 T1 Next
The T1 key is used to proceed to the next menu.
4.5.2 T2: Choose language
This key calls up the menu that allows you to select a language. The following choices are
available:
- German
- English
- French.
- T1 To make a selection
- T2 To confirm action and return to menu 1
4.5.3 T3: Ready
The T3 key is used to return to menu 1.
4.6 Menu 6
Transmissions
4.6.1 T1 Next
The T1 key is used to proceed to the next menu.
4.6.2 T2: Default values
The T2 key is used to display the LED transmissions:
- Red channel
output value: XXX
input value: XXX
- Green channel
output value: XXX
input value: XXX
The intensity of both channels (input values) must be above 500. If this value is not reached, the
device needs to be readjusted.
4.6.3 T3: Ready
The T3 key is used to return to menu 1.
5. Calibration
After starting, the device is calibrated. After prolonged operation, recalibration of the device is
recommended to compensate for any thermal drifts of the optical components, for example,
thereby ensuring accuracy of the dose evaluation.
5.1
Measurements with a dosimeter film WITHOUT supporting film
If this measurement method is used, ensure that there is no dosimeter film in the measuring
head during calibrating.
5.2
Measurements with both dosimeter film AND supporting film together
As the dosimeter film is normally supplied with a supporting film, measurements can also be
performed with supporting films. If this measurement method is used, calibration must be
performed with the supporting film only (i.e. without the dosimeter film).