Basic Setup and Operation
When both temperature points have been entered, the user may select the New
Curve field in order to generate the new curve. This will cause the display of a menu
like the one shown here:
CalGen
New Curve Menu
# User Sensor 1
Sets the curve number for the generated curve. Numeric entry. Note:
only the user curves can be written.
--Save--
Pressing the
Enter
key will cause the generation of a new curve. The
curve will be stored at the curve number specified on line 1.
Table 34: CalGen New Curve Menu
From this screen, the user must select the target user curve for the generated curve.
Finally, select the Save field in order to generate the curve and store it in the selected
user location.
Note: The CalGen process may be aborted by pressing the
Esc
or
Home
key.
Using CalGen With Platinum and Resistor Sensors
The calibration curve generation procedure for Platinum or Resistor sensors is the
same as for the diode sensors described above. However, these curves are
generated using two user specified points. Therefore, the selection of the number of
points is not required.
94