Cryo-con Model 24C
Remote Operation
RS-232 Configuration
The RS-232 interface supports baud rates of 9600, 19,200, 38,400, 57,600 and
115200. The factory default is 9600.
The baud rate is changeable from the instrument's front panel by using the
System
Functions Menu
.
Other RS-232 communications parameters are fixed in the instrument. They are set
as follows:
Parity: None
Bits: 8
Stop Bits: 1
Mode: Half Duplex
The RS-232 interface interface uses a "New Line", or Line Feed character as a line
termination. In LabView or the C programming language, this character is \n or
hexadecimal 0xA.
The controller will always return the \n character at the end of each line.
i
Note:
Some serial port software drivers allow the programmer
to set a line termination character. This character is then appended
to each string sent to the controller and stripped from returned
strings. In this case, the \n (0xA) character should be selected.
USB option configuration
Cryo-con's external USB option is automatically configured by the instrument when it
is plugged into the RS-232 port. Your computer will see it as an extra COM port. Use
it for communications just like any other RS-232 port.
104