Cryo-con Model 24C
Appendix E: Sensor Data
The BB package is easily mounted with a #4-40 brass screw. A brass screw is
recommended because thermal stress will be reduced at cryogenic temperature.
The mounting surface should be clean. A rinse with Isopropyl Alcohol is
recommended.
First, apply a small amount of Apiezon
f
N grease to the threads of the screw and on
the mounting surface of the sensor package.
Next, place the bobbin on the mounting surface, insert screw through bobbin and
lightly tighten.
The Canister Sensor Package
Cryo-con's Ruthenium-Oxide sensors are available in a small 0.95" x 0.2" cylindrical
canister package.
Construction:
Gold-plated cylindrical OHFC copper canister, Stycast
®
epoxy filler.
There is no internal atmosphere. Epoxy limits the maximum storage temperature
to 400K.
Leads:
Four, 36 AWG, Phosphor-Bronze, color coded. Formvar
®
insulation.
Mass:
0.4g.
Installation:
Use a 0.101" diameter drill. Place a small amount of Apiezon
®
N grease
in the hole before inserting the sensor. Ensure that the leads are thermally
anchored.
Connection:
All connections should be 4-wire in order to eliminate errors due to lead resistance.
Leads are coated with Butyl and may be separated by dipping them in Isopropyl
Alcohol.
Lead insulation is heavy Formvar® which is difficult to strip. Techniques include use of
a mechanical stripper or scrapping with a razor blade.
0.200in [5.08mm]
0.095in [2.41mm]
4 x AWG 36
Cable Color Code
V+
Clear
V-
Green
I+
Black
I-
Red
165