Cryo-con Model 24C
Appendix G: Sensor Data Tables
Appendix G: Sensor Data Tables
Silicon Diode
Silicon diode sensors offer good
sensitivity over a wide temperature range
and are reasonably interchangeable.
Use in magnetic fields is not
recommended.
Silicon diode sensors use a constant-
current DC excitation of 10
A.
Cryo-con S900 Silicon Diode
Name: Cryocon S900 Configuration: Diode
T(K)
Volts
mV/K
1.4
1.63864
-36.56
4.2
1.53960
-33.91
10
1.35568
-26.04
20
1.18193
-11.34
30
1.10465
-3.12
50
1.07188
-1.46
77.35
1.02511
-1.69
100
0.98615
-1.85
150
0.88988
-2.03
200
0.78311
-2.17
250
0.67124
-2.28
300
0.55674
-2.36
355
0.42759
-2.33
400
0.32161
-2.38
450
0.20231
-2.37
500
0.09077
-2.12
Cryo-con S800 Silicon Diode
Name: Cryocon S800 Configuration: Diode
T(K)
Volts
mV/K
1.4
1.87515
-36.86
4.2
1.75099
-49.16
10
1.47130
-43.45
20
1.18867
-15.93
30
1.10594
-3.90
50
1.07079
-1.47
77.35
1.02356
-1.86
100
0.98170
-1.85
150
0.88365
-2.03
200
0.77887
-2.13
250
0.67067
-2.20
300
0.55955
-2.22
355
0.44124
-2.10
385
0.37611
-2.26
Scientific Instruments SI-430 and SI-440
Name: SI 430 Diode Configuration: Diode
Name: SI 440 Diode Configuration: Diode
T(K)
Volts
mV/K
1.4
1.63864
-36.56
4.2
1.53960
-33.91
10
1.36317
-26.04
20
1.17370
-11.34
30
1.10343
-3.12
50
1.07399
-1.46
77.35
1.02511
-1.69
100
0.98740
-1.85
150
0.89011
-2.03
200
0.78272
-2.17
250
0.67085
-2.28
300
0.55665
-2.36
355
0.42759
-2.33
400
0.32161
-2.38
450
0.20231
-2.37
500
0.09077
-2.12
173