Cryo-con Model 24C
Specifications, Features and Functions
Resolution:
Shown below are typical RMS resistance noise values measured
at 50% of full-scale on a room-temperature resistor with a 3-Second analog
time-constant.
Range
10mV
3.0mV
1.0mV
40
W
1.0mA
255µ
W
1.0mA
255µ
W
100μA
2.6m
W
400
W
100μA
2.6m
W
100μA
2.6m
W
10μA
26m
W
4K
W
10μA
26m
W
10μA
26m
W
1.0μA
260m
W
40K
W
1.0μA
250m
W
1.0μA
250m
W
1.0μA
2.5
W
Table 8: Resolution for NTC Resistors
Accuracy:
Accuracy for the 10mV bias setting is specified in ranges
according to the following table. The formulas apply from the maximum to the
minimum resistance shown below.
Resistance
Range
Excitation
Current
Range
Min/Max
Resistance
Accuracy at 25C
1Ω
1.0mA
1 - 4Ω
±0.02Ω ± 0.05% * Rdg
4Ω
1.0mA
4 - 40Ω
±0.02Ω ± 0.05% * Rdg
40Ω
100µA
40 - 400Ω
±0.2Ω ± 0.05% * Rdg
400Ω
10µA
400 - 4KΩ
±2.0Ω ± 0.05% * Rdg
4KΩ
10µA
4K - 40KΩ
±20Ω ± 0.05% * Rdg
100KΩ
10µA
40K - 100KΩ
±50Ω ± 0.1% * Rdg
Table 9: 10mV Constant-Voltage Accuracy Specifications
While it is possible to measure resistance above100KΩ, accuracy is not
guaranteed.
The 1.0mV and lower bias settings are provided for use in very low
temperature applications (<~1K) where errors are often dominated by sensor
self heating rather than the accuracy of resistance measurement. In the
1.0mV range, the Model 24C will have an accuracy of ± 0.5% over the
resistance range of 40 to 10.0KΩ.