Cryo-con Model 24C
Specifications, Features and Functions
Measurement Resolution and Control Stability
The input analog-to-digital converter used by the Model 24C is 24 bits with no missing
codes. Thus, the measurement resolution is identifiable as one part in 2
-24
. However,
the only use for measurement resolution is to compute control stability. Since control
stability is limited by the output DAC rather than the input, the measurement
resolution specification is limited to one part in 2
-20
.
Where:
MR is the electronic measurement resolution in sensor units.
FullScale is the full scale range
MRT is the measurement resolution in temperature units.
SenSen is the sensor sensitivity at the measurement point.
29
MRT
MR
SenSen
MR
FullScale 2
20