Cryo-con Model 24C
Specifications, Features and Functions
Rear Panel
Figure 3: Model 24C Rear Panel Layout
AC Power Connection
The Model 24C requires single-phase AC power of 50 to 60 Hz. Voltages are set by
the line voltage selector in the Power Entry Module on the rear panel. The power cord
will be a standard detachable 3-prong type.
Line voltage selections are: 100, 120, 220 or 240VAC. Tolerance on voltages is +10%
to -5% for specified accuracy and -10% for reduced full-scale heater output in the
highest output range.
The power jack and mating plug of the power cable meet Underwriters Laboratories
(UL) and International Electrotechnical Commission (IEC) safety standards.
User-replaceable fuses are incorporated in the Power Entry Module.
i
Note:
The Model 24C uses a smart power on/off scheme.
When the power button on the front panel is pressed to turn the
unit off, the instrument's setup is copied to flash memory and
restored on the next power up. If the front panel button is not used
to toggle power to the instrument, the user should configure it and
cycle power from the front panel button one time. This will ensure
that the proper setup is restored when AC power is applied.
Caution
:
Protective Ground:
To minimize shock hazard, the
instrument is equipped with a three-conductor AC power cable. Plug
the power cable into an approved three-contact electrical outlet or
use a three-contact adapter with the grounding wire (green) firmly
connected to an electrical ground (safety ground) at the power outlet.
37