Cryo-con Model 24C
Specifications, Features and Functions
Fuse Replacement and Voltage Selection
Access to the Model 24C's fuses and voltage selector switch is made by using a
screwdriver to open fuse drawer in the power entry module. A slot is provided above
the voltage selector window for this purpose.
The fuse and voltage selection drawer cannot be opened while the AC power cord is
connected.
Voltage selection is performed by rotating the selector cams until the desired voltage
shows through the window shown.
There are two fuses that may be removed by pulling out the fuse modules below the
voltage selector. Fuses are specified according to the AC power line voltage used.
Mechanical, Form Factors and Environmental
Enclosure
The Model 24C enclosure is standard 2-U half-width 17-inch rack-mountable type that
may be used either stand-alone or incorporated in an instrument rack.
Dimensions are: 8.5"W x 3.5"H x 12"D. Weight is 9 Lbs.
An instrument bail and feet are standard. Rack Mount kits are available from Cryo-
con for both single instrument or side-by-side dual configurations. A rack mount kit is
optional.
Line Voltage
Fuse
Example
100VAC, 120VAC
2.0A slow-blow
Littlefuse 313 002
220VAC, 240VAC
1.0A slow-blow
Littlefuse 313 001
Table 14. AC Power Line Fuses
38