Cryo-con Model 24C
Front Panel Menu Operation
The System Configuration Menu
This menu is accessed by pressing the
System
key from the Home Status Display. It
is used to set many of the
instrument’s parameters
including display resolution,
I/O port settings etc.
System Configuration Menu
1
+Display TC: 4S
Sets the display time constant in seconds. Selections
range from 0.5S to 64S
2
+Display Res: 3
Sets the resolution. Selections are: 1, 2, 3 or Full.
3
Network Config
Press Enter to go to the network configuration menu.
4
+
Sets RS-232 port baud rate
5
+GPIB Adrs 12
Sets GPIB I/O address. (Note: GPIB is an external
option)
6
Datalog Config
Press
Enter
to go to the data logging setup screen.
7
Over Temp Config
Press
Enter
to go to the Over Temperature Disconnect
configuration screen.
8
FW Rev:1.01A
Displays the firmware revision level and hardware
revision letter.
9
+Pwr Up In Ctl:
Power Up Mode. Off for normal operation. On to engage
the control loops 10 seconds after power has been
turned on.
10
+AC Line: 60Hz
AC line frequency. Select 50 or 60Hz.
Table 20: System Configuration Menu
System Configuration Menu
Display TC: 4S
Display Res: 3
Network Config
RS232: 9600
GPIB Adrs: 12
Datalog Config
Over Temp Config
FW Rev: 1.00A
Pwr Up In Ctl: No
AC Line: 60Hz
Date: 08/01/2009
Time: 12:12:42
60