Cryo-con Model 24C
Introduction
Cryo-con's innovative instrument configuration menus show real-time status
information so the user can
instantly
view the results of any changes made.
On the control loop menu,
the controlling source
temperature, heater range
and power output level can
be observed while tuning a
loop.
An essential feature for
debugging system software
is the Network Configuration
Menu's ability to show
remote commands as they
are sent and received to the
instrument.
Sensor Curves:
The Model 24C includes built-in curves that support most
industry standard temperature sensors. Additionally, eight
user calibration curves
are available for custom or calibrated sensors. Each user curve may have up to 200
entries and are entered from the front panel, or transferred via any of the available
remote interfaces.
New calibration curves may be generated using the CalGen
feature to fit any existing
diode, Platinum or NTC resistor calibration curve at up to three user specified
temperature points. This provides an easy and effective method for obtaining higher
accuracy temperature measurements without expensive sensor calibrations.
Data logging
is performed by continuously recording to an internal 1,365 entry
circular buffer. Data is time stamped so that the actual time of an event can be
determined. Non-volatile memory is used so that data will survive a power failure.
Input Channel Statistics:
The Model 24C continuously tracks temperature
history independently on each input channel and provides a statistical summary that
indicates the channel's minimum, maximum, average and standard deviation. Also
shown are the slope and the offset of the best-fit straight line of temperature history
data.
Alarms:
Visual, remote and audible alarms are independently programmed to
assert, or clear based on high or low temperature condition, or a detected sensor
fault. Latched alarms are asserted on an alarm condition and will remain asserted
until cleared by the user.
Network Configuration Menu
Dev: M24C1234
DHCP Ena: On
Msk:255.255.255.0
00:50:C2:6F:40:3C
IP: 192.168.0.198
GWy:192.168.0.1
>input a:temp?;units?;name?;sys:time?
<0.5321;K;Sample Holder;14:37:25.
3
Loop 1A:Sample Holder
Set Pt:300.000K
Pgain: 6.000
Igain: 60.00S
Dgain: 7.500/S
Pman: 5.0000%
Type: RampT
Input: ChA
A: 123.456K
Ramp 42% of Mid
Range: MID
PID Table index: 2
Htr Load: 50
W
Next2