Cryo-con Model 24C
Basic Setup and Operation
Viewing a Cryocooler Thermal Signature
In order to view a cryocooler’s thermal signature and experiment with the
synchronous filter, the Cryo-con Utility Software may be used.
In the Data Logging menu, set the
interval
field to the minimum allowed value of 0.1
Seconds and then open a strip chart. Use the manual settings on the strip-chart to
zoom in on the temperature. The signature with the chart set to the base temperature
plus or minus about 0.5K should be observable.
In order to see the cooler signature, set the
Sync Filt. Taps
field to one. This disables
the removal of the signature. From here, you can enter various values in order to see
the affect of the synchronous filter.
Shown here is an example of a Cryomech PT403 pulse-tube refrigerator with a very
low heat-capacity load. The first part of the graph is with the synchronous filter turned
off and the second part shows a setting of 11 taps.
In most cases, a tap setting may be found that completely eliminates the signature.
90