Page 46
•
Crystal Engineering Corporation
Temperature.Module.(RTD100).Specifications
.
X
Accuracy
0% to 100% of Full Scale . . . . . . ±(0.015% of r 0.02Ω)
.
X
Ranges,.Resolutions.and.Units
Range
. . . . . . . . . . . .
0 to 400Ω for use with 100Ω PRT (platinum resistance temperature detectors)
Resolution
. . . . . . .
0.01 on all scales
Units
. . . . . . . . . . . . .
ºC, K, ºF, R, and Ω
.
X
Wiring.Types.and.TCRs
Wiring Types:
. . . .
2-, 3-, or 4-wire
Available TCR Selections:
0.00385, 0.003911, 0.003926
.
X
Connection
RTD Sensor Connector Interface:
P/N 3953: RTD Connection Kit (one IP67 terminal block connector)
(equivalent to Phoenix Contact Order Number: 1501265, or
Part Number:
SACC-M8MS-4CON-M-SM, or Binder P/N: 99-3383-100-04)
Cable Diameter . . . . . . . . . . . . . . . . . . . . . . . . 3.5 to 5.0mm [0.14 to 0.012 in] with smooth,
continuous covering adequate for IP67 sealing
Stranded Conductor Cross Section . . . . . 0.14 to 0.5mm
2
[0.0002 to 0.0008 in
2
]
.
X
Logging.Interval
Fastest Logging Interval
. . . . . . . . . . . . . . . . . . .
2 readings per second.
Although nVision logging interval may be set to a faster rate, the RTD100 module will update
the temperature reading at 2 readings per second or less.
.
X
Entity.Parameters
The RTD100 module has these specific input entity parameters:
Ui = 0
Uo = 9.73 V
li = 0
lo = 1.6642 V
Pi = 0
Po = 1.1 W
Co = 0.5 uF
Lo = 12 uH
*
*
Total cable inductance between all modules
Summary of Contents for nVision Lab Reference Recorder