Clamp-on amp sensor 0-1000A
Clamp-on amp sensor 0-1000A
English, V1.3, 05.01.11
English, V1.3, 05.01.11
0970 0033
0970 0033
2
3
Safety Specifications
Rated for 600 V dc/ac circuits of Overvoltage Category IV and 1000 V dc/ac
circuits of Overvoltage Category III per EN61010-1:2002 and EN 61010-2-032:2002.
Designed to Protection Class II, double or reinforced insulation requirements of UL
61010-1, 2nd Edition (2005.07), CAN/CSA-C22.2 No. 61010-1, 2nd Edition (2004.07)
and EN 61010-2-032:2002, 2nd Edition.
Current Ranges
0A to 1000A AC
Output Signals
4mA to 20mA DC
Accuracy (up to 400Hz)
±
(2% of rdg +5)
Working Voltage
600V rms
Common Mode Voltage
30 V AC
Safety Specifications
Storage Temperature
-20 ºC to 60 ºC (-4 ºF to 140 ºF)
Operating Temperature
0
ºC to 45 ºC (32 ºF to 113 ºF)
Altitude
0 to 2000 meters (0 to 6560 feet)
Relative Humidity
0% to 80% (0 ºC to 35 ºC; 32 ºF to 95 ºF)
0% to 70% (35 ºC to 45 ºC; 95 ºF to 113 ºF)
Maximum Conductor Size One 51 mm (2.0 inch)
Maximum Jaw Opening
52 mm (2.04 inch)
Dimensions
205 x 97 x 44 mm
(8.07 x 3.82 x 1.73 inch)
Weight
480g (16.9 oz.)
General Specifications
Operation Instructions
1
)
Connect the current sensor to one of DS 300 portable analog input terminals.
1) Center the single conductor inside the Probe jaws.
2) Make sure the Probe is perpendicular to the conductor.
3) Make sure the probe jaws are closed completely.
4) Avoid measurements, if possible, close to other current carrying conductors.
Guidelines for Positioning the Probe Jaws
Oder Information
Order No.
Description
0554 0504
Clamp-on amp sensor 0-1000A
2) Turn on DS 300 portable and set the corresponding terminal's input sensor type to
"Current".
3) Clamp the probe around a single conductor. Please refer to section "Guideline for
positioning the probe jams".
4
)
Read the measure value.
*
If “LOW BAT” LED on, please contact CS for service
DS 300-P6 analog input terminal 3 to 6
DS 300-P4 analog input terminal 3 to 4
Measuring Current