background image

 

 

Clamp-on amp sensor 0-1000A 

Clamp-on amp sensor 0-1000A 

English, V1.3, 05.01.11 

English, V1.3, 05.01.11 

0970 0033 

0970 0033 

Safety Specifications 

        

Rated for 600 V dc/ac circuits of Overvoltage Category IV and 1000 V dc/ac 

circuits of Overvoltage Category III per EN61010-1:2002 and EN 61010-2-032:2002.  

Designed to Protection Class II, double or reinforced insulation requirements of UL 
61010-1, 2nd Edition (2005.07), CAN/CSA-C22.2 No. 61010-1, 2nd Edition (2004.07)  
and EN 61010-2-032:2002, 2nd Edition.

 

Current Ranges 

0A to 1000A AC 

Output Signals 

4mA to 20mA DC 

Accuracy (up to 400Hz) 

±

(2% of rdg +5) 

Working Voltage 

600V rms 

Common Mode Voltage 

30 V AC 

Safety Specifications 

Storage Temperature 

-20 ºC to 60 ºC (-4 ºF to 140 ºF) 

Operating Temperature 

ºC to 45 ºC (32 ºF to 113 ºF) 

Altitude 

0 to 2000 meters (0 to 6560 feet) 

Relative Humidity 

0% to 80% (0 ºC to 35 ºC; 32 ºF to 95 ºF) 

0% to 70% (35 ºC to 45 ºC; 95 ºF to 113 ºF) 

Maximum Conductor Size   One 51 mm (2.0 inch)  

Maximum Jaw Opening 

52 mm (2.04 inch) 

Dimensions 

205 x 97 x 44 mm 

(8.07 x 3.82 x 1.73 inch) 

Weight 

480g (16.9 oz.) 

General Specifications 

Operation Instructions 

1

Connect the current sensor to one of DS 300 portable analog input terminals. 

1)   Center the single conductor inside the Probe jaws. 

2)   Make sure the Probe is perpendicular to the conductor.

 

3)   Make sure the probe jaws are closed completely.

 

4)   Avoid measurements, if possible, close to other current carrying conductors.

 

Guidelines for Positioning the Probe Jaws 

Oder Information 

Order No. 

Description 

0554 0504 

Clamp-on amp sensor 0-1000A 

2)  Turn on DS 300 portable and set the corresponding terminal's input sensor type to 
"Current". 

3)  Clamp the probe around a single conductor. Please refer to section "Guideline for 
positioning the probe jams". 

4

Read the measure value. 

If “LOW BAT” LED on, please contact CS for service

 

DS 300-P6 analog input terminal 3 to 6

 

DS 300-P4 analog input terminal 3 to 4

 

 

Measuring Current 

CS 

In

stru

me

nt

s (A

sia

) C

o.

,  L

td

.

Summary of Contents for 0554 0504

Page 1: ...heck surface of the Probe jaws these should be free of dust rust dirt and other foreign matter Never use a Probe whose insulating protection has been impaired Such probes should be made inoperative by taping the clamp shut in order to prevent operation CS provides a warranty for this product of 12 months covering the material and workmanship under the stated operating conditions from the date of d...

Page 2: ...0 to 70 35 ºC to 45 ºC 95 ºF to 113 ºF Maximum Conductor Size One 51 mm 2 0 inch Maximum Jaw Opening 52 mm 2 04 inch Dimensions 205 x 97 x 44 mm 8 07 x 3 82 x 1 73 inch Weight 480g 16 9 oz General Specifications Operation Instructions 1 Connect the current sensor to one of DS 300 portable analog input terminals 1 Center the single conductor inside the Probe jaws 2 Make sure the Probe is perpendicu...

Page 3: ...0 to 70 35 ºC to 45 ºC 95 ºF to 113 ºF Maximum Conductor Size One 51 mm 2 0 inch Maximum Jaw Opening 52 mm 2 04 inch Dimensions 205 x 97 x 44 mm 8 07 x 3 82 x 1 73 inch Weight 480g 16 9 oz General Specifications Operation Instructions 1 Connect the current sensor to one of DS 300 portable analog input terminals 1 Center the single conductor inside the Probe jaws 2 Make sure the Probe is perpendicu...

Page 4: ...heck surface of the Probe jaws these should be free of dust rust dirt and other foreign matter Never use a Probe whose insulating protection has been impaired Such probes should be made inoperative by taping the clamp shut in order to prevent operation CS provides a warranty for this product of 12 months covering the material and workmanship under the stated operating conditions from the date of d...

Reviews: