Installation / Connection / Measurement
DP400 mobil V1.00 EN
Seite 10 von 15
7.2
Measurement dew point
7.2.1 Measurement test set up/ test procedure
The maximum operating pressure is 15 bar.
It is important that the media connections are correctly plugged in and the
connection hoses are correctly installed.
Otherwise, the pressure tightness / strength of the device is compromised.
Before disconnecting the DP 400 mobile from the gas line, it is necessary to vent
the system to atmospheric pressure, otherwise serious injuries may occur.
Note:
At delivery, the dew point measuring device is designed in such a way that the sample gas is
introduced via the medium inlet (6 mm plug input) and with a direct flow out at the medium outlet, see
figure.
For applications of gases other than compressed air, it is mandatory to change the
media outflow (gas out) and to lead out it safely via a suitable line.
S“Gas Out“ direct flow out