Technical data DP 400 mobil  
 

 

DP 400 mobil  V1.00 EN 

 

Seite 6 von 15 

 

 

4  Technical Data DP 400 mobil 

 

 

Dimensions of housing 

270x225x156 mm 

Weight 

3,2 kg 

Case Material 

impact resistant HDPE/HWU 

–plastic (ABS). 

Sensor inputs 

 

Sensor inputs (digital / analogue)  depending on configuration 

Digital sensors RS 485/ModBus RTU 

Analogue third-party sensors 0/4 

– 20 mA, 0 - 1/10/30 V, pulse,              

Pt100/Pt1000. 

Measuring ranges 

 

Dew point:            -

80 … +50°Ctd 

Temperature: 

-

20 …70°C 

Humnidity: 

0 …100% RH 

Pressure: 

0 ...16 bar  

Accuracy dew point 

± 1°C for 20...-20°Ctd 

± 2°C for -20...-50°Ctd 

± 3°C for -50...-80°Ctd 

Accuray pressure 

± 0,5 % 

Power supply for sensors 

Output voltage:: 24 VDC ± 10% 

Output current  : a.) Digital board 120 mA continuous operation  

                           b.) Analogue board:  60mA  

Interfaces 

USB-Stick,  

Ethernet  RTU/TCP optional,  

WEB-Server optional 

Memory card 

Speichergröße 4 GB-Memorycard-Standard 

Power supply 

Li-Ion Battery,  

Power Supply 100 

– 240 VAC/50 – 60 Hz with output 12VDC 

Colour screen 

3.5“-Touchpanel TFT transmissive, graphics, curves, statistics 

Operating temperature 

– 50 °C 

Storage temperature 

-20 to +70°C 

Process connection 

6mm plug-in connection 

Optionally 

Ethernet Modbus TCP 

Optionally 

Webserver 

Optionally 

Mathematics calculation function 

 
 

Summary of Contents for DP 400 mobil

Page 1: ...DP 400 mobil V1 00 EN Seite 1 von 15 Installation and operating instructions portable dew point meter with integrated pressure measurement DP 400 mobil EN English...

Page 2: ...vour of the DP 400 mobil Please read this installation and operation manual carefully before mounting and initiating the device and follow our advice A riskless operation and a correct functioning of...

Page 3: ...nsors 7 6 Cable cross section 7 6 1 Sensor circuit points Output signal 7 7 Installation Connection Measurement 8 7 1 User controls connection 8 7 1 1 Dew point measurement 8 7 1 2 Display Datalogger...

Page 4: ...ons e g VDE 0100 Carry out maintenance only in strain less state All electric works are only allowed to be carried out by authorized qualified personnel Warning Inadmissible operating parameters Under...

Page 5: ...the Dangerous Goods Legislation requirements The user can transport the batteries by road without further requirements When being transported by third parties e g air transport or forwarding agency s...

Page 6: ...measured values of storage capacity Any analogue sensors 0 4 20 mA 0 1 10 30 V pulse are easy and fast to configure Digital sensors can be connected via RS 485 Modbus RTU Optional Network capable and...

Page 7: ...point 1 C for 20 20 Ctd 2 C for 20 50 Ctd 3 C for 50 80 Ctd Accuray pressure 0 5 Power supply for sensors Output voltage 24 VDC 10 Output current a Digital board 120 mA continuous operation b Analogu...

Page 8: ...y 0 03 mA 0 05 Input resistance 50 Voltage signal 0 1V Measuring range 0 1 V Resolution 0 05 mV Accuracy 0 2 mV 0 05 Input resistance 100 k Voltage signal 0 10 V 30 V Measuring range 0 10 V 30 V Resol...

Page 9: ...0 EN Seite 8 von 15 7 Installation Connection Measurement 7 1 User controls connection 7 1 1 Dew point measurement Plug in connection Medium closed with red seal plug Medium output Control Medium on o...

Page 10: ...ob during operation opens a popup with indication of the remaining operation time The data logger is operated via the touchpanel integrated in the display 7 1 3 Sensor connections Power supply connect...

Page 11: ...vice is compromised Before disconnecting the DP 400 mobile from the gas line it is necessary to vent the system to atmospheric pressure otherwise serious injuries may occur Note At delivery the dew po...

Page 12: ...ng point should be as close as possible to the critical measuring point in order to obtain a truly representative measurement The length of the DP 400 mobile cable should be as short as possible Only...

Page 13: ...ld be activated between the measurements or especially during longer periods of interruption and storage This is to protect the moisture sensor from excessively high moisture and thus to ensure fast r...

Page 14: ...sharp edges The foil can be damaged The DP 400 mobile is pre configured i e the pressure dew point temperature pressure as well as rel humidity are displayed directly on the start screen Configuratio...

Page 15: ...oap For decontamination spray the cleaner on an unused cotton cloth or one way wipe and wipe the component comprehensively Perform the final drying with a clean cloth or by air drying In addition the...

Reviews: