- 15 - 

 

 

The operation in detail 

 
By means of the "Enter" key the menu can be accessed. DS 300 mobile version has 
the following main menus: 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Logger operation 

 
 
DS 300 includes a data logger. In order to configure the logger function and to 
activate/deactivate it there are several functions available: 

 

Measuring rate 

Set logging rate and averaging option

  

The logging rate defines the interval data should be recorded in 
the memory. It will record from every activated channel a 
sample. The averaging option can be used to calculate an 
average value. I.e. DS 300 is measuring every 1 second and if 
the logging rate is 10 seconds, it will calculate an average value 
out of the last 10 measurement values and store it as a 
recorded sample. 

Manual start 

Start/stop logging 

Starts or stops the data recording. Whenever a new recording is 
started a new file is created in the memory. 

Select logging  
channels 

Select those channels which should be recorded in a file. 
Channels which are not selected are not stored. The selection 
is effected by placing a checkmark. 

Set logger memory  
mode 

There are following possibilities: 
1. "Wrap aound mode" - if the memory is full the first measured 
    values will be overwritten. 
2. "Stop if the memory is full" 

Show logger  
memory 

Shows the status and the size of the available memory 

 
 

 
Logger operation 
 
Sensor settings 
 
System status and settings 
 
                    Page 1 of 2 

Back 

Enter 

Summary of Contents for DS 300 mobile version

Page 1: ...V4_03_2011 Instruction manual D DS S 3 30 00 0 m mo ob bi il le e v ve er rs si io on n Multi channel display and data logger...

Page 2: ...er You have made the right decision by choosing a measuring instrument from CS Instruments GmbH Thousands of customers buy our high standard products every year There are a few good reasons for doing...

Page 3: ...ours Up to 6 parameters can be recorded at the same time consumption pressure dew point current consumption pressure temperature Special features All relevant measured data at a glance Actual air cons...

Page 4: ...charging time 10 hours Display Graphic display 160 x 100 pixels with back light Accuracy Dew point see sensor specifications Consumption see sensor specification Pressure 0 5 F S Current 5 25 A 2 100...

Page 5: ...y instructions Observe measuring ranges of the sensors Observe the admissible storage and transportation temperature as well as the permitted operating temperature e g protect the instrument from dire...

Page 6: ...tting 20 C 1000 hPa All volume flow values m h and consumption values indicated in the display are related to 20 C 1000 hPa according to ISO 1217 intake condition 0 C and 1013 hPa standard cubic meter...

Page 7: ...ange from to Measuring range from to 1 4 6 DN 6 1 0 157 l min 2 0 190 l min 1 2 16 1 DN 15 3 5 1516 l min 6 0 1836 l min 3 4 21 7 DN 20 0 4 178 m h 0 7 215 m h 1 27 3 DN 25 0 6 295 m h 1 1 357 m h 1 1...

Page 8: ...50 bar standard Accuracy 0 5 Ctd 10 to 50 Ctd typical 2 Ctd at 40 Ctd Operation temperature 20 to 70 C Storage temperature 40 to 80 C Screw in thread G1 2 stainless steel 3 measured value per FA 410...

Page 9: ...imum 4 hours without any mains supply In case of plugged in mains supply the rechargeable batteries will be loaded also if the instrument is switched off 1 2 3 Sensor connections factory settings 1 Co...

Page 10: ...10 10 Survey operation menu with Enter one menu level down and return with Back...

Page 11: ...1 USB connection icon This icon shows that DS 300 mobile version is connected to a PC via USB 2 Logger module status icon Logger module status WAIT Time start conditions set waiting for the sart of lo...

Page 12: ...nsor setting System status and setting Item 2 of 5 Back Enter Selectable menu item it shows 4 items at a time If the total items in the current menu level is more than 4 items the menu should scroll o...

Page 13: ...the e key to go back to the last menu level without activating the selected option Item 1 of 2 Back Enter Example Time and date setting 1 Use S and T key to adjust the time 2 Use f key to move from H...

Page 14: ...lly and start to display real time measurement values acquired from those sensors The measurement values may be displayed on more than one page To see another page just press the or button Pipe 1 flow...

Page 15: ...e logging rate is 10 seconds it will calculate an average value out of the last 10 measurement values and store it as a recorded sample Manual start Start stop logging Starts or stops the data recordi...

Page 16: ...can be made Set inner diameter For calculation of volume flow consumption and flow Set total consumption counter Counter can be re set to zero or set to any value Enter volume flow unit Selection of...

Page 17: ...recognized so the analogue sensor for the respective channel has to be selected The analogue inputs are preconfigured ex factory however they can be freely adjusted via the keypad For this purpose the...

Page 18: ...ration software Attention Maximum current supply 12 VDC 25 mA If the non CS analogue sensors have a higher current consumption please use an own power supply Connection cables CS Instruments offers co...

Page 19: ...ses or if the serial number in the instrument has been changed damaged or removed The warranty time for DS 300 is 12 months If no other definitions are given the accessory parts have a warranty time o...

Page 20: ...94 1886 Precision pressure sensor CS 40 0 40 bar accuracy 1 0694 0356 Precision pressure sensor CS 1 6 absolute 0 1 6 bar absolute accuracy 1 0694 3551 Clamp on ammeter 0 1000 A AC incl 5 m connection...

Page 21: ...21 Notes...

Page 22: ...22 Notes...

Page 23: ...23 Notes...

Page 24: ...4955 Harrislee Phone 49 0 461 700 2025 Fax 49 0 461 700 2026 info cs instruments com www cs instruments com Sales office South Germany CS Instruments GmbH Zindelsteiner Stra e 15 D 78052 VS Tannheim P...

Reviews: