- 16 -
Time start
Set time, start conditions
DS 300 can be programmed to start at a certain time.
Show protocol
Single protocols (files) can be deleted or the whole memory can
be freed. DS 300 will show the available protocols with date,
number of channels and number of recorded values per
channel.
Delete protocol
Deletes the selected protocol
Format logger
Deletes the internal logger
11.3 Sensor settings
11.3.1 Consumption and dew point sensors
The consumpiton and dew point sensors of CS Instruments are automatically recognized
at channel 1 or 2.
11.3.1.1 Setting of consumption sensor VA 400
(Attention: Please adjust inner diameter of the pipe!)
For CS consumption sensors the following settings can be made:
Set inner diameter
: For calculation of volume flow, consumption and flow
Set total consumption counter
: Counter can be re-set to zero or set to any value
Enter volume flow unit
: Selection of the desired flow unit
Adjust consumption unit
: The consumption unit is fixed by selecting the volume
flow unit
Adjust reference pressure
: In order to calculate the standard volume flow *
Adjust reference temperature
: In order to calculate the standard volume flow *
Adjust gas type
: Adjust the gas type in which measurement is carried out
*
Please observe
:
The reference pressure and the reference temperature are not referred to the current
process pressure or process temperature. They are used in order to calculate the standard
volume flow in case of standard conditions, e. g. 1000 hPa, 20 °C.
Changes of the sensor settings are immediately stored in the sensor as soon as they are
confirmed with <Enter>.