background image

 

 

- 17 - 

 

 

11.3.1.2  Setting of dew point sensor FA 410 

 
In this menu the dew point unit can be selected. The following units are preset ex factory: 
Dew point in °Ctd, temperature in °C and relative humdity in % RH 
 

ƒ

 

Setting of the humitidy unit

: (ºCtd, g/m3, g/kg, ppm etc.) Attention: ppm (V/V) and g/kg 

require the input of a reference pressure (absolute line pressure) 

ƒ

 

Setting of the reference pressure

: Required for ppm (V/V) and g/kg (absolute line 

pressure) 

ƒ

 For 

the 

atmospheric dew point

 please set the absolute line pressure. In case of this 

measuring unit the dew point under atmospheric conditions is calculated by means of the 
absolute line pressure. 

 
 

11.3.2  CS Analogue sensors (pressure sensor, clamp-on ammeter,  
            temperature sensor, …) 

 
DS 300 mobile version has 2 resp. 4 analogue input channels. The analogue input channels 
are not automatically recognized so the analogue sensor for the respective channel has to be 
selected. The analogue inputs are preconfigured ex factory, however they can be freely 
adjusted via the keypad. 
 
For this purpose the menu point "sensor setting", cannel 3 to 6 has to be selected. The 
following options are available: 
 

"Select sensor"

:  

In case of DS 300 mobile version there is a list of 6 sensors  

 

ready for selection. Please choose the required sensor.  

 
 

In order to complete the list with customer-specific sensors  

 

these customer-specific sensors can be defined by means of  

 

the software "DS 300-P configuration tool" and transferred 

 

to DS 300 mobile. 

 

"Show settings"

In this point the sensor-specific data are indicated. 

 

"Calibration offset"

Here possible zero-point errors of sensor and DS 300 can be  

 

adjusted. Typical example: One point calibration of a pressure  

 

sensor . The adjustment value stays stored in DS 300 in the  

 respective 

measuring 

channel. 

 

"Delete calibration offset"

:  Here the measured value entered under "calibration offset" can 

 be 

deleted. 

 

 
 
 

Summary of Contents for DS 300 mobile version

Page 1: ...V4_03_2011 Instruction manual D DS S 3 30 00 0 m mo ob bi il le e v ve er rs si io on n Multi channel display and data logger...

Page 2: ...er You have made the right decision by choosing a measuring instrument from CS Instruments GmbH Thousands of customers buy our high standard products every year There are a few good reasons for doing...

Page 3: ...ours Up to 6 parameters can be recorded at the same time consumption pressure dew point current consumption pressure temperature Special features All relevant measured data at a glance Actual air cons...

Page 4: ...charging time 10 hours Display Graphic display 160 x 100 pixels with back light Accuracy Dew point see sensor specifications Consumption see sensor specification Pressure 0 5 F S Current 5 25 A 2 100...

Page 5: ...y instructions Observe measuring ranges of the sensors Observe the admissible storage and transportation temperature as well as the permitted operating temperature e g protect the instrument from dire...

Page 6: ...tting 20 C 1000 hPa All volume flow values m h and consumption values indicated in the display are related to 20 C 1000 hPa according to ISO 1217 intake condition 0 C and 1013 hPa standard cubic meter...

Page 7: ...ange from to Measuring range from to 1 4 6 DN 6 1 0 157 l min 2 0 190 l min 1 2 16 1 DN 15 3 5 1516 l min 6 0 1836 l min 3 4 21 7 DN 20 0 4 178 m h 0 7 215 m h 1 27 3 DN 25 0 6 295 m h 1 1 357 m h 1 1...

Page 8: ...50 bar standard Accuracy 0 5 Ctd 10 to 50 Ctd typical 2 Ctd at 40 Ctd Operation temperature 20 to 70 C Storage temperature 40 to 80 C Screw in thread G1 2 stainless steel 3 measured value per FA 410...

Page 9: ...imum 4 hours without any mains supply In case of plugged in mains supply the rechargeable batteries will be loaded also if the instrument is switched off 1 2 3 Sensor connections factory settings 1 Co...

Page 10: ...10 10 Survey operation menu with Enter one menu level down and return with Back...

Page 11: ...1 USB connection icon This icon shows that DS 300 mobile version is connected to a PC via USB 2 Logger module status icon Logger module status WAIT Time start conditions set waiting for the sart of lo...

Page 12: ...nsor setting System status and setting Item 2 of 5 Back Enter Selectable menu item it shows 4 items at a time If the total items in the current menu level is more than 4 items the menu should scroll o...

Page 13: ...the e key to go back to the last menu level without activating the selected option Item 1 of 2 Back Enter Example Time and date setting 1 Use S and T key to adjust the time 2 Use f key to move from H...

Page 14: ...lly and start to display real time measurement values acquired from those sensors The measurement values may be displayed on more than one page To see another page just press the or button Pipe 1 flow...

Page 15: ...e logging rate is 10 seconds it will calculate an average value out of the last 10 measurement values and store it as a recorded sample Manual start Start stop logging Starts or stops the data recordi...

Page 16: ...can be made Set inner diameter For calculation of volume flow consumption and flow Set total consumption counter Counter can be re set to zero or set to any value Enter volume flow unit Selection of...

Page 17: ...recognized so the analogue sensor for the respective channel has to be selected The analogue inputs are preconfigured ex factory however they can be freely adjusted via the keypad For this purpose the...

Page 18: ...ration software Attention Maximum current supply 12 VDC 25 mA If the non CS analogue sensors have a higher current consumption please use an own power supply Connection cables CS Instruments offers co...

Page 19: ...ses or if the serial number in the instrument has been changed damaged or removed The warranty time for DS 300 is 12 months If no other definitions are given the accessory parts have a warranty time o...

Page 20: ...94 1886 Precision pressure sensor CS 40 0 40 bar accuracy 1 0694 0356 Precision pressure sensor CS 1 6 absolute 0 1 6 bar absolute accuracy 1 0694 3551 Clamp on ammeter 0 1000 A AC incl 5 m connection...

Page 21: ...21 Notes...

Page 22: ...22 Notes...

Page 23: ...23 Notes...

Page 24: ...4955 Harrislee Phone 49 0 461 700 2025 Fax 49 0 461 700 2026 info cs instruments com www cs instruments com Sales office South Germany CS Instruments GmbH Zindelsteiner Stra e 15 D 78052 VS Tannheim P...

Reviews: