- 18 -
11.3.3 Connection of non-CS analogue sensors
Own analogue sensors 0/4…20 mA, 0…1/10V, Pt100, Pt1000 can be connected to the input
channels 3, 4, 5 and 6. For this purpose the specific sensor datea have to be stored in DS
300 by means of the DS 300 mobile version configuration software.
Attention
:
Maximum current supply 12 VDC / 25 mA! If the non-CS analogue sensors
have a higher current consumption please use an own power supply!
Connection cables:
CS Instruments offers connection cables with open ends. Order no. 0553 0110, cable length:
5 m.
Connection scheme:
Attention:
DS 300 mobile version is configured on our premises. If you connect your
own analogue sensors DS 300 has to be configured newly by means of the
DS 300 configuration software
.