background image

 

 

- 5 - 

 

 
3.  Key features of DS 300 mobile version 

 

  Graphic display for easy user interface 

  Flexible power supply: 100…240 V AC/ 50…60Hz  resp. internal rechargeable battery 

  2 sensor inputs for CS consumption and dew point sensors (sensor connection 1 + 2) 

 USB 

interface 

  Data logger function for up to 1,000,000 values 

  PC software for easy configuration and evaluation of the data 

  Analogue input module (0…20 mA, 0…10 V, Pt100, Pt1000) (sensor connection 3, 4, 5 

and 6) 

 
 

4.  Calibration/adjustment 

 

At CS Instruments 

According to DIN ISO certification of the measuring instruments we recommend to calibrate 
and if applicable to adjust the instruments regularly at the manufacturer. The calibration 
intervals should comply with your internal specificaton. According to DIN ISO we recommend 
a calibration interval of one year for the instrument DS 300 including probes. 
 
 

5.  Safety instructions 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Observe measuring ranges of the sensors! 
 
Observe the admissible storage and transportation temperature as well as the 
permitted operating temperature (e. g. protect the instrument from direct 
insolation). 
 
The manufacturer cannot be held liable for any damage which occurs as a 
result of non-observance or non-compliance with these instructions. Should the 
device be tampered with in any matter other than a procedure which is 
described and specified in the manual, the warranty is cancelled and the 
manufacturer is exempt from liability. 
 
CS Instruments GmbH offers no gurantee for the suitability for any other 
purpose and is not liable for errors which may have slipped into this operation 
manual. CS Instruments GmbH is also not liable for consequential damage 
resulting from the delivery, capability or use of this device. 
 
We offer you to take back the instruments DS 300, which you would like to 
dispose of. 
 
Adjustments and calibrations should only be carried out by qualified employees 
from the measurement and control technology branch. 

Summary of Contents for DS 300 mobile version

Page 1: ...V4_03_2011 Instruction manual D DS S 3 30 00 0 m mo ob bi il le e v ve er rs si io on n Multi channel display and data logger...

Page 2: ...er You have made the right decision by choosing a measuring instrument from CS Instruments GmbH Thousands of customers buy our high standard products every year There are a few good reasons for doing...

Page 3: ...ours Up to 6 parameters can be recorded at the same time consumption pressure dew point current consumption pressure temperature Special features All relevant measured data at a glance Actual air cons...

Page 4: ...charging time 10 hours Display Graphic display 160 x 100 pixels with back light Accuracy Dew point see sensor specifications Consumption see sensor specification Pressure 0 5 F S Current 5 25 A 2 100...

Page 5: ...y instructions Observe measuring ranges of the sensors Observe the admissible storage and transportation temperature as well as the permitted operating temperature e g protect the instrument from dire...

Page 6: ...tting 20 C 1000 hPa All volume flow values m h and consumption values indicated in the display are related to 20 C 1000 hPa according to ISO 1217 intake condition 0 C and 1013 hPa standard cubic meter...

Page 7: ...ange from to Measuring range from to 1 4 6 DN 6 1 0 157 l min 2 0 190 l min 1 2 16 1 DN 15 3 5 1516 l min 6 0 1836 l min 3 4 21 7 DN 20 0 4 178 m h 0 7 215 m h 1 27 3 DN 25 0 6 295 m h 1 1 357 m h 1 1...

Page 8: ...50 bar standard Accuracy 0 5 Ctd 10 to 50 Ctd typical 2 Ctd at 40 Ctd Operation temperature 20 to 70 C Storage temperature 40 to 80 C Screw in thread G1 2 stainless steel 3 measured value per FA 410...

Page 9: ...imum 4 hours without any mains supply In case of plugged in mains supply the rechargeable batteries will be loaded also if the instrument is switched off 1 2 3 Sensor connections factory settings 1 Co...

Page 10: ...10 10 Survey operation menu with Enter one menu level down and return with Back...

Page 11: ...1 USB connection icon This icon shows that DS 300 mobile version is connected to a PC via USB 2 Logger module status icon Logger module status WAIT Time start conditions set waiting for the sart of lo...

Page 12: ...nsor setting System status and setting Item 2 of 5 Back Enter Selectable menu item it shows 4 items at a time If the total items in the current menu level is more than 4 items the menu should scroll o...

Page 13: ...the e key to go back to the last menu level without activating the selected option Item 1 of 2 Back Enter Example Time and date setting 1 Use S and T key to adjust the time 2 Use f key to move from H...

Page 14: ...lly and start to display real time measurement values acquired from those sensors The measurement values may be displayed on more than one page To see another page just press the or button Pipe 1 flow...

Page 15: ...e logging rate is 10 seconds it will calculate an average value out of the last 10 measurement values and store it as a recorded sample Manual start Start stop logging Starts or stops the data recordi...

Page 16: ...can be made Set inner diameter For calculation of volume flow consumption and flow Set total consumption counter Counter can be re set to zero or set to any value Enter volume flow unit Selection of...

Page 17: ...recognized so the analogue sensor for the respective channel has to be selected The analogue inputs are preconfigured ex factory however they can be freely adjusted via the keypad For this purpose the...

Page 18: ...ration software Attention Maximum current supply 12 VDC 25 mA If the non CS analogue sensors have a higher current consumption please use an own power supply Connection cables CS Instruments offers co...

Page 19: ...ses or if the serial number in the instrument has been changed damaged or removed The warranty time for DS 300 is 12 months If no other definitions are given the accessory parts have a warranty time o...

Page 20: ...94 1886 Precision pressure sensor CS 40 0 40 bar accuracy 1 0694 0356 Precision pressure sensor CS 1 6 absolute 0 1 6 bar absolute accuracy 1 0694 3551 Clamp on ammeter 0 1000 A AC incl 5 m connection...

Page 21: ...21 Notes...

Page 22: ...22 Notes...

Page 23: ...23 Notes...

Page 24: ...4955 Harrislee Phone 49 0 461 700 2025 Fax 49 0 461 700 2026 info cs instruments com www cs instruments com Sales office South Germany CS Instruments GmbH Zindelsteiner Stra e 15 D 78052 VS Tannheim P...

Reviews: