7. DS 300 mobile version with FA 410 dew point sensor
Initiation
1. Connect dew point sensor FA 410 to channel 1 or 2
2. Switch on the instrument (DS 300 will recognize the dew point sensor FA 410
automatically)
3. The following parameters will be indicated: dew point in °Ctd, temperature in °C,
relative humidity in % RH
7.1 Technical data dew point sensor FA 410
Measuring range
-80 to 20 °Ctd
Pressure range
-1 to 50 bar standard
Accuracy
±
0.5 °Ctd (-10 to 50 °Ctd)
typical
±
2 °Ctd at -40 °Ctd
Operation temperature
-20 to 70 °C
Storage temperature
-40 to 80 °C
Screw-in thread
G1/2" stainless steel
3 measured value per
FA 410
Dew point in °Ctd, temperature in °C, relative humidity in
% RH
Please read carefully before starting the device!
1. Attention: Do not exceed pressure range of > 50 bar with standard version.
(Up to 350 bar in case of special version).
2. Important: Before installation briefly bleed the compressed air in order to remove
condensate and particles. This prevents soiling of FA 410. Standing air leads to
long measuring times.