Operation
Page 34
FA 550 V1.00
Settings
4-20mA
Channel 1
Error Current
12.3.8 FA 550 Info
Settings
Info
This determines what is output in case of an error at the
analog output.
2 mA Sensor error / System error
22 mA Sensor error / System error
None Output according Namur (
3.8mA
– 20.5 mA)
< 4mA to 3.8 mA Measuring range under range
>20mA to 20.5 mA Measuring range exceeding
To make changes first select a menu item "Current Error"
with button
„
“
and then select by pressing the
„OK“
the desired mode
With
„back“
change to the settings menu.
Here you get a brief description of the sensor data incl.
some production data