Connection diagrams of the different sensor types
PI 500
Seite
11
von
93
6
Connection diagrams of the different sensor types
6.1
Connector pin assignment for all sensors at PI 500
The interface connector to be used is a ODU Medi Snap 8 pin
– Reference: K11M07-P08LFD0-6550
.
Available connection cables at CS-Instruments are:
ODU with Open ends:
Order no 0553 0501, cable length: 5 m.
Order no 0553 0502, cable length: 10 m.
ODU with M12 Connector:
Order no 0553 0503, cable length: 5 m.
Extension cable (ODU/ODU): Order no 0553 0504, cable length: 10 m.
Connection scheme:
Connector housing
Plug insert
Collet
Clamping nut
View on welding pins of Medi Snap Connector
Cable lenght 5m/10 m
1
weiss white
2
braun brown
3
grün
green
4
gelb
yellow
5
grau
grey
6
rosa
pink
7
blau
blue
8
rot
red
I (500µA)
1
2
3
4
5
6
7
8
+ RS485
- RS485
SDI
Analog IN +
+VB 24Vdc
-VB GND
Analog IN -
+ RS485
- RS485
SDI
(CS-internal data transmission for all Dew point and Flow sensor FA/ VA 400)
ANALOG IN +
ANALOG IN
–
)
STROMQUELLE 500 µA
+VB, 24V DC Power supply for sensor
-VB, GND Sensor
White
Brown
Green
Yellow
Grey
Pink
Blue
Red