Sensor-Settings / Configuration of Analogue -Sensors
PI 500
Seite
45
von
93
7.3.2.1.2.10 Configuration of Analogue-Sensors
A brief overview of the possible
Type
of settings with examples.
For
CS-Digital
7.3.2.1.2.2 Choice of the sensor type (For example type CS-Digital sensor)
and
7.3.2.1.2.6 Dew Point sensor with type CS-Digital
and
7.3.2.1.2.7 Consumption Sensor of type CS-Digital (SDI
The
Alarm-settings,
Record
-
Button, the
Resolution
of the decimal places and
Short Name
and
Value-Name
are all described in Chapter
7.3.2.1.2.11 Type 0 - 1/10/30 Volt and 0/4
– 20 mA
Main menu
Settings
Sensor settings
C1
Type description field
0 - 1/10/30 V
125. 44 °C
Main menu
Settings
Sensor settings
C1
arrow right (2.page)
Please see the scale of the sensor (here for
example
Type
0 - 10V
corresponds to
0 - 250 ° C) from the data sheet of the
connected sensor.
By
Scale 0V
enter the lower and
by
Scale10V
the upper scale value.
By
Scale 0V
enter the lower and
by
Scale10V
the upper scale value
The
Sensor Supply Voltage
is switched
On
,
if it´s required by the sensor type, otherwise
off (no green hook).
Please confirm by pressing the
OK
button
It is possible to define a Offset-Value. With
the
Set Value to-
button
(Offset
) you enter it.
The positive or negative difference of the
Offset
will be displayed.
By pressing the
Reset
-button the
Offset
will
be deleted