Data logger settings
PI 500
Seite
55
von
93
7.3.2.1.3 Data logger Settings
Main menu
Settings
Logger settings
Remark:
If more than 12 measurement data are recorded at the same time, the smallest possible time
interval of the data logger is 2 seconds.
In addition, if more than 25 measurement data are recorded at the same time, the smallest
possible time interval of the data logger is 5 seconds.
In the top row you can select the predefined
Time intervals
1, 2, 5, 10, 15, 30, 60 and 120
seconds for recording.
A different, individual
Time interval
can be
entered in the highlighted white description
field right at the head, where the currently set
Time interval
is always displayed.
Remark:
The largest possible
Time interval
is 300
seconds.