background image

0970 0043 Ex Documentation VA 450/452 English, V1.4, 05.11.12 

0970 0043 Ex Documentation VA 450/452 English, V1.4, 05.11.12 

EX Documentation VA 450/452

EX Documentation VA 450/452 

To all terminals of the flow transmitter, only devices with ratings Um 

 24 V DC and Im 

 500 mA are allowed to be connected. 

The flow transmitter is only allowed to be used in the permitted tem-
perature class.  

For ambient temperatures below –20 °C cables, drain  plugs, cable 
glands and cable entries must be suitable and certified. 

Cables and pipe entries must be certified (Ex d IIC) and must be 
suitable for operating temperatures of up to 105 °C . When using pipe 
entries, the associated sealing equipment must be fitted directly at 
the housing. 

The cable entries and openings not used must be sealed tight with 
suitable components. 

The rotation of the display is only allowed during installation and in 
non explosive environment! Before rotating  the local display make 
sure that the device is powered off for at least 5 minutes. 

The display version of VA 452 can be rotated in both directions 
270° . For this purpose the nut at the top of the s haft has to be 
opened fully. Then pull up the metal casing and rotate it in the de-
sired direction. It can be rotated in 90° steps in  both directions

metal pin locks the position. When the desired position is achieved, 
press down the metal casing onto the shaft so that the metal pin can 
insert into the position hole. Tighten up the nut. 

 

Important 
notes 

2. Installation

 

 

 

 

Examination 
certificate, 
certificates of 
conformity 

The system meets the fundamental health and safety requirements 
for the design and construction of devices and protective systems 
intended for use in potentially explosive atmospheres in accordance 
with appendix II of directive 94/9/EC. 

Other tech-
nical infor-
mation 

Since this manual only contains ex-relevant information please refer
to the instruction manual of VA 450 and VA 452 for technical data 
and further information. 

Front and 
back cover
security seal

The front and back cover of the instrument are secured with a screw 
to avoid the opening of the covers through unauthorized persons.
Please ensure that both security screws are tighten up after finishing
the installation! 

Security screws at back and front cover

Used 

standards 

EN 60079-0:2009

EN 60079-1:2007

Reviews: