Impedo
DUO
User Guide
29
Figure 7: Left side of instrument
Figure 6 show the right side of the instrument with USB port 1 and the GPS antenna
connector. Figure 7 shows the left side of the instrument with USB port 2 and the
two Ethernet communication ports.
2.4.2
Power Supply
The Impedo Duo can be powered and charged via a power factor corrected universal
input mains supply or via IEEE 802.3-2008 compliant Power over Ethernet (PoE). A
built in LiFePO
4
battery will power the instrument for up to ±20min in the absence of
power. This battery has a design life expectancy of more than 2,000 charge/discharge
cycles.
The Power Supply port contains 3 terminals on a removable lockable plug. Each
terminal can accommodate conductors with size up to 2.5mm
2
.
The supply voltage input range is from 90 V
RMS
to 300 V
RMS
AC or DC (42-69Hz for AC
sources). This enables the instrument to be powered from the 110 V VT terminals,
230 V auxiliary supply or the DC bus.