Impedo
DUO
User Guide
37
2.4.6
Relay Outputs
Six relay outputs (100 mA
RMS
, 350 V
RMS
) are used to provide pulsed outputs or to
control external processes. The outputs are arranged into 2 galvanically isolated
clusters (1 kV isolation); each containing 3 switched contacts with a common return
path.
The relay outputs are available on an 8-terminal lockable connector plug with screw
terminals. Each terminal can accommodate conductors with size up to 1.5mm
2
.
Table 5: Relay switch connection layout
Terminal
Legend
Description
1.
S
1
Switch A1
2.
S
2
Switch A2
3.
S
3
Switch A3
4.
C
m
Switch A Common
5.
S
4
Switch B1
6.
S
5
Switch B2
7.
S
6
Switch B3
8.
C
m
Switch B Common
Figure 14: Relay switch connection layout
Sw
it
ch
B
C
o
m
m
o
n
Sw
it
ch
B
3
Sw
it
ch
B
2
Sw
it
ch
B
1
Sw
it
ch
A
C
o
m
m
o
n
Sw
it
ch
A
3
Sw
it
ch
A
2
Sw
it
ch
A
1
Summary of Contents for Impedo DUO
Page 85: ...Impedo DUO User Guide 85 Figure 49 Export harmonic selection...
Page 109: ...Impedo DUO User Guide 109 Appendices 8 Impedo DUO dimensions...
Page 110: ...Impedo DUO User Guide 110 8 1 Panel mounting plates 8 2 19 installation plate...
Page 111: ...Impedo DUO User Guide 111 9 Impedo DUO wiring diagram...
Page 113: ...Impedo DUO User Guide 113 11 Portable ImpedoDUO...
Page 117: ...Impedo DUO User Guide 117 Screw driver USB with meter key Ethernet cable...