2.4.8
Ethernet Ports
The instrument has two Gigabit Ethernet ports as shown in Figure 16. The IP address
of port 1 is fixed at 192.168.7.1. The instrument serves as a DHCP server for the
network connected to this port. The IP address of port 2 can be set using the GUI
(see Section 0) and is also used to interface to GPRS/3G routers for remote
communication.
Do not connect Ethernet port 1 to a network with an active
DHCP server, as this may cause the network to stop functioning
properly
The IP address of Ethernet port 1 is fixed at 192.168.7.1
The instrument serves as a DHCP server for equipment
connected to port 1
IEEE 802.3-2008 compliant Power over Ethernet (PoE) is available on both Ethernet
ports. The PoE capability enables the Impedo Duo to be powered through the
communication infrastructure. The Impedo Duo will therefore be available at all
times while the communication infrastructure stays on-line.