SECTION 3: OPERATION
M N X 1 0 0 0 1 , R E V C
●
2 / 8 / 2 0 1 2
9
Connection Technology Center, Inc.
●
7939 Rae Boulevard
●
Victor, NY 14564
●
(585) 924-5900
Operating Procedure
To operate the Signal conditioner, make sure that all wires are properly
connected, then apply power.
CAUTION: Make sure that power input does not exceed specified limits or
damage to the system may result.
Make sure the green Sensor OK light is illuminated.
If using BNC Signal Output interface with a portable data collector that supplies
constant current power, turn the power supply off in the data collector prior to
collecting data.
Analog Output Zero Adjustment
This section contains test setup and procedures for adjusting the zero. The
zero adjustment has been factory set and normally requires no adjustments. If
an adjustment becomes necessary, it can be accomplished as described below.
For a current output during setup, monitor the output with a DC current meter
connected to the output. A minimum series load resistor of 50 ohms should be
used (maximum load resistor is 1 K). For the voltage output, monitor the output
with a DC voltmeter.
1.
Remove transducer input.
2.
Adjust the ZERO ADJUST for a 4.00 mA reading.