Calibration
Calibrating the Ultrasound Sensors
CubiScan 125
55
4. The following screen is displayed. Place the calibration cube on the
platform and slide it against the back corner until it is in contact with
both side panels with the 12'' side against the left side panel (when
looking from the front) as shown below. Tap
[Next]
to continue.
Figure 53
Third Sensor Calibration Screen