Troubleshooting
Diagnostics
CubiScan 125
71
2. Tap
DIAGNOSE
.
Select the Gate option if it is not already
selected.
From this screen you can view the sensitivity threshold board settings.
The three LED beam bars now shown on the display represent the LED
beams that the CubiScan 125 uses to measure objects. This is a useful
screen for determining the functionality of the LED beams.
The width and height bars correspond to the width and height boards in the
measurement gate.
These LED beams bars are blue when no light beams are being broken (as
seen above), which typically means that there is no object in the
measurement field.
The emitter checkbox should always be checked when you are measuring
objects. When this box is unchecked, the gate is not emitting light and
cannot measure objects.
Figure 67
First Gate Diagnostic Screen