SPECIFICATIONS
PARAMETER
VALUE
Power Requirements
Power:
100VAC / 0.12A maximum
240VAC / 0.05A maximum
Frequency: 50/60Hz
Video Outputs
2x HDMI, 1x DVI
Console Dimensions
11.2”W x 4.5”L x 13.2”T
Console Weight
4.6 lbs.
Camera Head Weight
0.85 lbs.
Transport & Storage
Conditions
-
40 to 122⁰F (
-
40 to 50⁰C)
10 to 90%, non-condensing
Operating Conditions
50 to 86⁰F (10 to 30⁰C)
30 to 75% RH, non-condensing
1.4.
CLASSIFICATIONS
PARAMETER
VALUE
System Classification
FDA Class I
EU Class I
Health Canada Class I
Design Criteria
IEC 60601-1 Ed. 3.1: 2012
EMC Certifications
CISPR 11 Class A
IEC 60601-1-2 Ed. 4
CE Marking
CE Marking for MDD 93/42/EEC
Protection against electric
shock.
Class I [grounded]
Degree of protection against
harmful ingress of water
Camera Console is [IPX-0]; no protection.
Degree of safety in the
presence of Flammable
Anesthetics
Equipment is NOT suitable for use in the presence of
flammable anesthetics.
Mode of operation
Continuous