3500 - ActiveCare Console
UM-040
Optional Accessories
(Sold Seperately)
3501 ActiveCare Pager, 2588 Personal Emergency Transmitter
(PET), 3502 Wireless Handset, 2552 DoorAlert Strip Monitor,
2558 Wireless Call Button with Pull Cord, 2559 Wireless Reset
Button, 2572 Wristband Transmitter, 2575 Caregiver Reset
Transmitter, 2642 Caregiver Reset Key, 3503 Wireless Magnetic
Reed Switch, 2662 Cordless Bed Pad, 2661 Cordless Chair
Pad, 3515 Cordless Under Mattress Pressure Pad, 2663
Cordless Floor Mat, 2719 Cordless Pad Transmitter, 2404 Plug
Pack 12Vdc 2A
Wall mountable receiver/transmitter with a large two-digit
LED display. Supports pad, call, and door type alarms,
and signals to pager.
Volume Control
Button Lock
- Disables the “Program”,
“Select 1” and “Select 2” buttons to prevent
accidental programming.
Tracking
- Turn on to monitor for lost pad
transmitters.
Power Adaptor Socket
Accepts the 2404 12V Power Adaptor.
Cable Strain Relief
2719 pad transmitter activation procedure:
To activate a 2719 transmitter please do the following:
1. Connect the transmitter to a chair / bed pad or
floor mat.
2. Turn ON the transmitter.
3. For bed and chair pads press and hold the pad for
15 seconds
4. For floor mats stand on the mat for 15 seconds.
The transmitter is now activated.
Please note:
If the transmitter is turned OFF at any stage, then
the 15 second activation procedure will have to be performed
again.
Select 1
Select 2
Review
Reset
“Tens” Digit
and
the
“Units” digit
Power Indicator LED
Panel Battery Indicator LED
Pad Indicator LED
Call Indicator LED
Sensor / Door Indicator LED
Signal Lost
Device Battery
Program
Alarm tone
There are two alarm tones that can be selected. To change the
alarm tone place the Console into programming mode (see
Programming Mode
instruction on this page).
Now press the “Program” button again until the tens digit
displays a “b”. Then using the “Select 2” button to select either
(tone) 1 or (tone) 2.
Area code
If several Consoles are being used in a facility, then setting an
Area Code will ensure pagers only receive alarms from their
specified area. Please refer to the full documentation or your
cura1 distributor for instructions on how to do this.
Pause setting
The Console can be programmed to allow staff to pause all pad
alarms for a designated time period. Please consult the full
documentation or your distributor if you would like to enable this
feature.
Delay Setting
The Console can be programmed to automatically delay a pad
alarm for a designated time period. This Please consult the full
documentation or your distributor if you would like to enable this
feature.
Programming Mode
To put the Console into programming mode, press and
hold the “
Program
” Button (located on the front of the
Console) for approximately 5 seconds. The Console will
initiate a long beep and briefly display “PP” to confirm
that it is in programming mode. Now release the
programming button. The front display will display the
first memory slot “01”. This will always be the default
view when entering programming mode.
Call forwarding to a pager or slave console
The Console can be configured to forward calls (alarms) to the
3501 ActiveCare Pagers and to additional 3500 ActiveCare
Consoles that have been configured as Slaves.
To enable call forwarding place the Console into programming
mode (see
Programming Mode
instruction on this page).
Now press the “
Program
” button again until the tens digit
displays a “t”. Then using the “
Select 2
” button to select the
setting from the table below:
Set t =
Call Forward Setting
0
Forwarding OFF
1
Not used
2
Not used
3
Forwards to 3501 ActiveCare Pager and Slave
Console
4
Console is configured as a Slave