background image

1

1

1

QUICK START GUIDE

3010 SeizureAlert

Step    

1

Step    

2

Connecting a 3

rd

 Party Alarm Device

Note - if you are using only the 3501 ActiveCare Pager as the 

alarming device, then this step is not required.

Plug the 6.25 mm plug on the Alarm Lead (14) into your alarm 

device (eg nurse call system, emergency dialer, audible or 

visual alarm). 
For 3

rd

 party systems that use a plug size other than 6.25mm, 

cut off the 6.25mm plug and replace it with a compatible plug or 

connector. 
Plug the other end of the Alarm Lead into the back of 3010 

SeizureAlert Monitor where it is labelled "Out". 

4.2 MICROPHONE Sensitivity Adjustment:

 It is advisable to only 

use the microphone if prolonged regular groans or verbal sounds 

normally accompany the seizure, otherwise leave the 

MICROPHONE control set to the “Min” position. 
Short duration sounds will not activate the alarm. Adjust the 

MICROPHONE control so that each time you make a loud groan 

sound (like clearing your throat) the DETECT LED illuminates. As 

with the BED SENSOR, the alarm will only activate after ~3 - 15 

seconds of detected sounds. (Exact alarm activation time will 

depend on the DELAY setting in the next step.)

4.3 DELAY Adjustment: 

If detecting a seizure quickly is 

imperative, then little or no delay interval should be used however 

a restless sleeper may cause the alarm to activate by simply 

moving around in the bed.
To adjust the delay interval first set the DELAY to the "Min" position 

and tap the bed until the monitor alarms. After ~10 seconds the unit 

will reset & the red alarm LEDs will go out. Increase the DELAY 

setting by rotating the control clockwise a little, and test again. 

Repeat this process until you are happy with the delay setting.

Note - the delay setting also affects the alarm response time of the 

microphone (if in use).

4.4 Save Settings and Exit Setup Mode: 

Once the required 

sensitivity has been achieved, press and hold the red "Setup" 

button until the red and orange LEDs flash simultaneously. This 

indicates that the settings have been saved by the monitor. 

Operation

To reset the unit after an alarm, simply turn it off by 

pressing the ON / OFF button. 

Sensor

(Under Mattress)

OPTION

MENU

CANCEL

Sensor

Out

Setup

240 V/AC

Nurse Call 

or 3

rd

 Party 

Alarm

Back of 3010 

SeizureAlert

3501 

ActiveCare 

Pager (3011 only)

Position the Monitor and Sensor

Place the monitor on a bedside table or other position where it 

is unlikely to be accidentally damaged.
Position the Bed Sensor between the base of the bed and the 

middle of the mattress.
Plug the Power Adaptor into 240 V/AC outlet & switch on. 

Switch the 240 V/AC outlet on but leave the monitor OFF. 

4.1 BED SENSOR Sensitivity Adjustment: 

Turn the BED 

SENSOR control to the 12 o’clock position and the MICROPHONE 

and DELAY controls to the “Min” position (anticlockwise).
Tap the centre of the bed with your hand while observing the 

DETECT LED, which will flash GREEN when the BED SENSOR 

control is set high enough to detect the movement. If the DETECT 

LED does not illuminate then turn the BED SENSOR control 

incrementally higher until the unit detects the tapping on the bed. 
Continue to tap the bed with quick tapping actions for at least 5 - 10 

seconds, or until the POWER and DETECT LEDs both turn solid 

red. A connected alarm device should now be alarming.
For most applications the 12 o’clock position will be adequate.

Please read all of steps 1 - 5 before beginning.

1.  Make sure the monitor is plugged into mains power and 

is turned on at the mains but turned OFF at the monitor.

2. Turn the DELAY control to maximum.
3. (While performing the following step, please watch the 

POWER LED and DETECT LED.) Hold down the red 

Setup button on the back of the unit while turning the 

unit ON. Keep the Setup button held in. 

4. After a short period the LEDs on the front panel will 

flash alternately GREEN / ORANGE five times to 

indicate that the monitor is now in setup mode. You can 

now release the Setup button.

5. Proceed to 

Step 4.1 - 4.3

 to adjust the settings for BED 

SENSOR, MICROPHONE and DELAY. When testing 

the settings the monitor alarms are indicated by the 

POWER & DETECT LEDs both going RED, then after 

approx. 5 seconds the monitor will return to setup mode 

for additional sensitivity adjustment, if required.  

6.  To save settings and exit setup mode, press and hold 

the red Setup button until the POWER & DETECT LEDs 

flash 5 times and go off. 

Enter Setup Mode

Step    

3

Configure the 3010 SeizureAlert Monitor

During setup, the monitor settings can be reset at any time by 

simply switching the monitor off using the ON / OFF button. The 

monitor will exit setup mode and revert to the settings that were 

saved before entering setup mode. 
Before switching on again please wait 5 - 10 seconds until the 

POWER and MOVEMENT LEDs are no longer illuminated. 
It is not neccesary to have a 3

rd

 party alarm connected during 

the setup process but be sure to test it after setup. 

Step    

4

Note - during normal operation the controls at the front of the 

monitor will have no effect on the unit. This is to prevent accidental 

adjustments to the settings. 