MHW 0002 – DK 1693 | Version 001 – 14/09/2017 | custo med GmbH
2
[2] Resting and Stress ECG hardware, custo cardio 200 |
page 12
Resting and Stress ECG
Hardware, description of device for custo cardio 200
2.5
Preparing the patient and examination procedure
Prerequisites for an inspection:
Proper installation, configuration and commissioning of the system.
Resting ECG
Make sure that the ECG device is connected to the PC
and the power supply to the device is guaranteed.
Check that your patient is lying comfortably and is not cold.
Start custo diagnostic and click:
Examination, Resting ECG, New Resting ECG see
Software description...
.
Spray the contact spray
1)
on the electrode application points.
Place the electrodes on the patient according to the placement diagram,
See
2.6 Attaching the device to the patient, electrode application
.
Start the recording.
The patient must keep still during the recording process.
Stress test ECG
Make sure that the ECG device is connected to the PC
and the power supply to the device is guaranteed.
When using an ergometer, make sure that the patient is in the optimal
seating position (the extended leg should be slightly bent).
When using a treadmill, make sure to follow the instructions
in
2.7 Instructions for stress ECG with treadmill
.
Start custo diagnostic and click:
Examination, Stress ECG, New Stress ECG see
Software description...
.
Spray the contact spray
1)
on the electrode application points.
Place the electrodes on the patient according to the placement diagram,
See
2.6 Attaching the device to the patient, electrode application
.
Wait some minutes so that the contact between the skin
and the electrodes can develop optimally.
Apply the blood pressure cuff.
Start the recording.
1) Contact spray warning
Before each examination, make sure to spray contact spray on the electrode
application points. Never spray the contact spray onto the electrodes or into the
electrodes or leads. Do not use water or contact gel!
Recommended contact sprays from:
C+V Pharma Depot GmbH or GE Medical Systems
Only use recommended contact sprays.
Residues from other products can damage the custo cardio 400
electrodes.