background image

MHW 0012 – DK 1713  |  Version 004 – 20/04/2020  |  custo med GmbH

2

Resting and Stress ECG

Hardware, description of device for custo cardio 400 BT and custo cardio 400 accu

[2] Resting and Stress ECG Hardware, custo cardio 400 devices | 

page 17

2.5.5 

custo cardio 400 troubleshooting (both device variants)

In the event of an error, the custo cardio 400 suction level display lights up yellow 

or red. The cause of a warning is saved in the custo cardio 400 error log file and 

can be viewed there.

In order to open the error log file, the custo cardio 400 device must be connected to 

the PC via USB. In this case, the device is displayed as custo cardio 400 mass stor-

age. Open the directory of the custo cardio 400 mass storage via Windows Explorer. 

The file is called „error.log“ and can be opened with a text editor.

  Warning indicator lights up yellow

The system is leaking

The device tries to gain the preset pressure for over five minutes, e.g. 280 mbar 

(suction level 6). If this does not succeed, the device switches off and the warning 

indicator lights up yellow.

Troubleshooting: 

Start the device without a patient (press the On/Off button). If 

the device controls extremely often (engine noise), there is a mechanical problem:

   one or more sealing rings on the leads are missing / stuck in the housing

   the electrodes are not tight (dirt, e.g. from hair)

   the device has an internal defect (service case)

If the device is tight without a patient,

check and correct the following points when used on patients:

   Selection of the suitable electrode type (small, normal, hair)

   Placing the electrodes on a less hairy part of   skin / shave patient

   Selection of a lower suction level which is more easily achieved.

    Note: electrodes adhere less strongly, possibly reduced quality

    of the derivations.

The pressure in the system is too high

The system may be clogged, e.g. too much moisture in the leads. Start a cleaning 

cycle with the cleaning button (longer blowing out of the leads).

If this error still occurs or if the red light indicator 

 

appears during cleaning, con-

tact your authorised custo med dealer.

  Warning indicator lights up red

Defect on the device

The device must be sent in for repair. Contact your authorised custo med dealer.

Summary of Contents for custo cardio 400 accu

Page 1: ...ECG Part 2 Hardware description of device for custo cardio 400 BT and custo cardio 400 accu 1 2 3 4 Safety Hardware Software Hygiene Operating characteristics 12 channel PC ECG Resting stress ECG MHW...

Page 2: ...nslated into another language without the prior written consent of custo med GmbH The manufacturer reserves the right to change the information in this Operating Manual without prior notice The curren...

Page 3: ...sto cardio 400 display and operating elements 14 2 5 3 custo cardio 400 device states in standard operation 15 2 5 4 custo cardio 400 further states and special functions 16 2 5 5 custo cardio 400 tro...

Page 4: ...rer custo med GmbH Maria Merian Str 6 85521 Ottobrunn Germany Production date YYYY MM e g 2018 01 CE mark notified body Serial number Follow the Operating Manual Medical electrical equipment classific...

Page 5: ...used to attach the electrodes to the patient s body using underpressure R L and F according to Einthoven and C1 to C6 according to Wilson are used as derivations Intracardiac application is not intend...

Page 6: ...plication system 6 thoracic wall leads 1 20 m and 4 extremity leads 1 45 m Electrode hair other versions optional Spacer 2 fold Spacer 3 fold Telescopic boom for custo cardio 400 BT Bracket for telesc...

Page 7: ...s 1 20 m and 4 extremity leads 1 45 m Electrode hair other versions optional Spacer 2 fold Spacer 3 fold Extension arm custo move for custo cardio 400 accu incl USB cable Fixture for the extension arm...

Page 8: ...on swivel range 270 Continuous height adjustment with gas pressure spring Gas pressure spring Extraction lock to extend and pull in the boom arm Bracket for custo cardio 400 BT with safety bar Adjust...

Page 9: ...USB cable to the USB port of the custo cardio 400 BT device route it along the boom and fix it with the Velcro fasteners IMPORTANT Do not connect the USB cable to the PC until after the software inst...

Page 10: ...usto move must be ordered se parately custo move may only be mounted in the new fixture from 2020 with radial groove see Screw the fixture onto the equipment cart with the groove facing backwards Use...

Page 11: ...to the left and turn custo move to the desired position To move custo move safely from one place to anoth er align the extension arm straight forward The an ti rotation lock is located centrally on t...

Page 12: ...the coding labels to the ends of the patient leads The labels C1 to C6 are glued to the chest wall leads length 1 20 m The labels R L F and N are glued to the longer extremity leads length 1 45 m Conn...

Page 13: ...ice is then in standby mode On Off button is constantly illuminated in blue When inactive the device is put into energy saving mode after one minute the On Off button flashes blue and the battery char...

Page 14: ...for manually blowing out the lines States of the suction level and status display During operation when performing an ECG recording the suction level and status display provides information about the...

Page 15: ...CG interface is called up in custo diagnostic the suction level control is activated and the electrode leads can be applied to the patient The patient s ECG signal is displayed on the screen In this s...

Page 16: ...wer in custo diagnostic must be at least set at level 3 With the automatic suction level control the suction level is increased until the system is tight or the electrodes are sufficiently tight again...

Page 17: ...hes off and the warning indicator lights up yellow Troubleshooting Start the device without a patient press the On Off button If the device controls extremely often engine noise there is a mechanical...

Page 18: ...y of the device is guaranteed The battery charge should be 20 charge status indicator blue or green not purple Charge the device if necessary or use the mains When using an ergometer make sure that th...

Page 19: ...rib between C2 and C4 V4 C4 brown 5th intercostal space on the left midclavicular line V5 C5 black on the left hand side of the anterior axillary line at the same height as C4 V6 C6 purple on the left...

Page 20: ...der to avoid excessive movement artefacts and therefore interference in the other terminal lines During the ECG recording the electrode leads must not come into contact with the patient the treadmill...

Page 21: ...supply Battery information Charging time approx 4 5 hours at 1 5 A charging current in standy longer charging time during operation Running time approx 9 hours at suction level 3 varies depending on...

Page 22: ...s For proper operation it is necessary to use the operating system software combinations tested and approved by custo med for the respective custo diagnostic version also custo diagnostic server and c...

Page 23: ...k connection not for single position systems Fanless Dual DVI or DP graphics card for CPET Windows 8 1 x64 PRO version for joining a domain Ports One USB 2 0 port per USB device preferably not USB 3 0...

Page 24: ...ltage fluctuations flickers according to IEC61000 3 3 Compliance Group 1 Class B Class A Complies Immunity tests Static electricity discharge ESD according to IEC 61000 4 2 Quick transient electric in...

Page 25: ...z and 40 66 MHz to 40 70 MHz The amateur radio bands between 0 15 MHz and 80 MHz are 1 8 MHz to 2 0 MHz 3 5 MHz to 4 0 MHz 5 3 MHz to 5 4 MHz 7 MHz to 7 3 MHz 10 1 MHz to 10 15 MHz 14 MHz to 14 2 MHz...

Page 26: ...e directly next to other devices or stacked together with other devices should be avoided as this could result in fault operation If the devices must nonetheless be used as described above this device...

Page 27: ...04 2020 custo med GmbH 2 Resting and Stress ECG Hardware description of device for custo cardio 400 BT and custo cardio 400 accu 2 Resting and Stress ECG Hardware custo cardio 400 devices page 27 2 1...

Page 28: ...13 Version 004 20 04 2020 custo med GmbH 2 Resting and Stress ECG Hardware description of device for custo cardio 400 BT and custo cardio 400 accu 2 Resting and Stress ECG Hardware custo cardio 400 de...

Page 29: ...Bluetooth 4 0 USB Adapter 16018 1 USB extension cable 1 8 m plug A socket A 85023 Power supply unit BET 0612 for custo cardio 400 12 V 5 A 85044 Power supply cable 1 m for power supply unit BET 0612...

Page 30: ...and hex key 85081 Telescopic boom short for equipment cart custo easy plus 85070 Cable hook for telescopic boom 85073 Bracket for telescopic boom incl mounting material 85082 Extension arm custo move...

Page 31: ...Manufacturer s contact details custo med GmbH Maria Merian Stra e 6 85521 Ottobrunn Germany Phone 49 0 89 710 98 00 Fax 49 0 89 710 98 10 E mail info customed de Internet www customed de...

Reviews: